D.O. CALIBRATION
15
y
When the reading is stable, “READY” & “ACCEPT” tags start blinking. Press ACCEPT to
confirm the “100.0%” D.O. calibration.
y
If the reading is within the limits (±15% f.s.), the meter stores the value (and adjusts
the slope point).
y
The instrument stores the slope calibration data and returns to measurement mode.
Notes:
If the reading is not close to the selected value, “WRONG“ & “BUFFER” tags will blink.
MW190 has automatic buffer recognition function. If the ARROW keys are pressed
to select the desired calibration value, the automatic buffer recognition function is
disabled.