Dear Customer,

Thank you for choosing a Milwaukee Instruments 

product. This manual will provide you with the necessary 

information for the correct use of the instrument. Please 

read it carefully before using the meter. 

SPECIFICATIONS

Range

0.00 to 50.00 FNU

50 to 1000 FNU

Resolution

0.01 FNU and 1 FNU

Precision

±0.5FNU or ±5% of reading, whichever 

is greater

Light Source

High emission infrared LED

Light Detector

Silicon Photocell

Method

Detection of scattered light, according 

to ISO 7027

Environment

0 to 50 °C (32 to 122 °F) 

100% RH max

Battery Type

1 x 9 volt

Auto-Shut off

After 5’ of non-use

Dimensions

192 x 104 x 52 mm 

(7.5 x 4.1 x 2”)

Weight

380 g

MEASUREMENT PROCEDURE

1.  Turn the meter on by pressing ON/OFF.

2.  When the LCD displays “- - -”, the 

meter is ready.

3.  Fill a clean, dry cuvet with sample up 

to the mark.

4.  Replace the cap.
5.  Clean the cuvet with a lintfree cloth 

to remove any fingerprints, dirt or 

waterspots.

6.  Place the cuvet into the holder and 

ensure that the notch on the cap is 

positioned securely into the groove.

7.  Press READ/▲ and “SIP” will blink 

during measurement.

 

8.  The instrument directly displays turbidity in FNU units.

CALIBRATION PROCEDURE

1.  Turn the meter on by pressing ON/OFF. When the LCD 

displays “- - -”, the meter is ready.

2.  Press CAL key and “CAL” will blink on the display.

 

3.  Press CAL key again. The date of calibration in MM.DD 

format appears on the LCD.

 

4.  To toggle between month and day 

press DATE/▶ key.

5.  To change the date press READ/▲ 

key.

6.  To confirm the date values press CAL 

key.

A blinking ZERO message will appear on the LCD. 

Insert the 0 FNU standard and press CAL key. “SIP” 

will blink during measurement.

 

▶ 

 ▶ 

7.  At the end the blinking “10.00” message will appear 

on the LCD.

Insert the 10 FNU standard and press CAL key. “SIP” 

will blink during measurement.

8.  The blinking “500” message will appear on the LCD.

Note:

 At this moment you can exit calibration by 

pressing READ/ key and the meter will memorize only 

the two-point calibration.

Insert the 500 FNU standard and press CAL key. “SIP” 

will blink during measurement

9.  At the end of the reading the instrument goes directly 

to measurement mode. Now the instrument is 

calibrated and ready to use.

Note:

 A monthly calibration is recommended.

The instrument can be checked with the supplied 

standard solutions each day.

To check the date of the last calibration simply press 

DATE/▶ key.

GUIDE TO DISPLAY CODES

This prompt appears for 1 second each 

time the instrument is turned on.

The dashes “- - -” indicates that 

the meter is in a ready state and 

measurement can be performed.
Sampling In Progress. Flashing “SIP” 

prompt appears each time the meter 

is performing a measurement.
The presence of battery icon on the 

display indicates that the battery 

voltage is getting low and the battery 

needs to be replaced.
“-BA-”, the battery is dead and must 

be replaced. Once this indication is 

displayed, the meter will lock up. Change 

the battery and restart the meter.
“CAL”, the meter is in calibration mode. 

If the CAL key is not pressed within 6 

seconds, the meter will automatically 

switch to idle mode.
“CL” appears on the lower part

of the display when the meter

is in calibration mode.