Dear Customer,

Thank you for choosing a Milwaukee Instruments 

product. This manual will provide you with the necessary 

information for the correct use of the instrument. Please 

read it carefully before using the meter. 

SPECIFICATIONS

Range

0.00 to 5.00 mg/L Cl

2

Resolution

0.01 mg/L (0.00 - 3.50 mg/L) 

0.10 mg/L (3.50 mg/L)

Precision

±0.06 mg/L @1.50 mg/L

Light Source

Tungsten lamp

Light Detector

Silicon Photocell and 525 nm narrow 

band interference filter

Method

Adaptation of the USEPA method 330.5 

and Standard Method 4500-Cl G

Environment

0 to 50 °C (32 to 122 °F) 

100% RH max

Battery Type

9 volt (1 pc)

Auto-Shut off

After 10’ of non-use

Dimensions

192 x 104 x 52 mm 

(7.5 x 4.1 x 2”)

Weight

380 g

MEASUREMENT PROCEDURE

1.  Turn the meter on by pressing ON/OFF.

2.  When the LCD displays “- - -”, the 

meter is ready.

3.  Fill the cuvet with 10 mL of unreacted 

sample, up to the mark, and replace 

the cap.

4.  Place the cuvet into the holder and 

ensure that the notch on the cap is 

positioned securely into the groove.

5.  Press ZERO and “SIP” will blink on the 

display.

6.  After a few seconds the display will 

show “-0.0-”. The meter is now zeroed 

and ready for measurement.

 

 

7.  Remove the cuvet.

8.  For free chlorine measurement: add 3 drops of DPD 1 

indicator and 3 drops of DPD 2 buffer to an empty 

cuvet. Immediately add 10 mL of unreacted sample. 

Replace the cap and swirl gently.

10 ml

9.  Immediately reinsert the cuvet into the holder 

and ensure that the notch on the cap is positioned 

securely into the groove.

 

10 mL

 ▶

DPD 1

DPD 2

x 3

x 3

+

10. Press READ and “SIP” will blink during measurement.

11.  The instrument directly displays concentration in 

mg/L of free chlorine.

12. Then, for total chlorine measurement, remove the 

cuvet and add to the reacted sample 3 drops of 

DPD 3. Replace the cap and swirl gently.

DPD 3

x 3

13. Reinsert the cuvet into the holder and ensure that the 

notch on the cap is positioned securely into the groove.

14. Wait for 2 minutes and 30 seconds then press READ 

and “SIP” will blink during measurement.

2’30”

15. The instrument directly displays concentration in 

mg/L of total chlorine.

Interferences

 

y

Bromine (positive error)

 

y

Chlorine dioxide (positive error)

 

y

Iodine (positive error)

 

y

Oxidized Manganese (positive error)

 

y

Chromium (positive error)

 

y

Ozone (positive error)

 

y

Alkalinity above 250 mg/L CaCO

3

 or acidity above 150 

mg/L CaCO

3

 will not reliably develop the full amount of 

color or it may rapidly fade. To resolve this, neutralize 

the sample with diluted HCl or NaOH.

 

y

In case of water with hardness greater than 500 mg/L 

CaCO

3

, shake the sample for approximately 2’  after 

adding the reagent.

GUIDE TO DISPLAY CODES

This prompt appears for 1 second 

each time the instrument is turned 

on.
The dashes “- - -” indicates that the 

meter is in a ready state and zeroing 

can be performed.
Sampling In Progress. Flashing “SIP” 

prompt appears each time the meter 

is performing a measurement.

“-0.0-”, the meter is in a zeroed state 

and measurement can be performed.

The blinking “BAT” indicates that the 

battery voltage is getting low and the 

battery needs to be replaced.
“-bA-”, the battery is dead and must 

be replaced. Once this indication 

is displayed, the meter will lock up. 

Change the battery and restart.
“Conf”, the meter has lost its 

configuration. Contact your dealer 

or the nearest Milwaukee Customer 

Service Center.

ERROR MESSAGES

On Zero Reading

 

y

Blinking “-0.0-” indicates that the 

zeroing procedure failed due to a 

low signal-to-noise ratio. In this 

case press ZERO again.

 

y

“no L”, the instrument cannot 

adjust the light level. Please check 

that the sample does not contain 

any debris.

 

y

“L Lo”, there is not enough light to 

perform a measurement. Please 

check the preparation of the zero 

cuvet.

 

y

“L Hi”, there is too much light to 

perform a measurement. Please 

check the preparation of the zero 

cuvet.