background image

EC/TDS OPERATIONS 

• The user can select the EC/TDS 
conversion  factor (CONV), as well as 
the ß coefficient  (BETA) for automatic 
temperature compensation. 
• Enter the EC or TDS mode with the 
“RANGE”  key, then press and hold the 
“ON/OFF” key until “TEMP” is displayed 
on the secondary 
LCD. Press this key 
again to  show  the 
current conversion 
factor (CONV). 
• Press the “SET” 
button to change  the 
value, then press 
again this key to show 
the current 
temperature coefficient (BETA). 
• Press “SET” to change the value, then 
press “ON/OFF” key to return to normal 
operation. 
• For taking EC or TDS measurements, 
immerse  the probe in the solution to be 
tested. 
• Select either EC or TDS mode with 
“RANGE”. 
• Stir gently and wait for the reading to 
stabilize 

(the clock symbol stops 

blinking). 
 

EC/TDS CALIBRATION PROCEDURE 

• Clean the probe with 

M10000 

solution. 

• Enter the EC mode with the “RANGE” 
key. 
• Immerse the probe in 

MA9030 

(12.88 

mS/cm) calibration solution. 
• Press and hold the 
“ON/OFF” key until 
“CAL”  is displayed on 
the LCD. Release  the 
button and the 
message “12.88 mS USE” appears. 

• The meter will perform automatic 
calibration, and  then display “OK” for 1 
second before returning  to normal 
mode. 
• The calibration is now complete and 
the “CAL” tag is displayed on the LCD to 
indicate that the meter is calibrated. 
 

BATTERY REPLACEMENT 

The meter shows the remaining battery 
percentage every time it is switched on. 
When the battery level is below 5%, the 
battery symbol on the bottom left  of the 
LCD lights up to alert the user. 
The meter is also provided with BEPS 
(Battery  Error Prevention System), 
which automatically  switches the meter 
off if the battery level is so low to cause 
erroneous readings. It is recommended 
to replace immediately the battery. 
Turn the meter off, 
remove the  battery 
compartment cover 
from  the rear of the 
meter and replace 
the rundown 9V 
battery with a  new 
one. 
Install the battery 
while paying 
attention to its 
polarity and reattach 
the cover. 
 

NOTES 

• If reading is over-range, the display will 
flash the closest full scale value. 
• To clear a previous calibration, press 
“ON/  OFF” after entering the calibration 
mode. The  lower LCD will display ESC 
for 1 second and the meter will return to 
normal measurement  mode. The “CAL” 
symbol on the LCD will  disappear and 

the meter is reset to the default 
calibration. 

 
ACCESSORIES 
MA851D/1 

Amplified 

pH/EC/TDS/ 

Temperature  probe with 
DIN connector and 1 m 
(3.3’) cable. 

MA9004  

pH 4.01 buffer, 230 ml bottle 

MA9006  

pH 6.86 buffer, 230 ml bottle 

MA9007  

pH 7.01 buffer, 230 ml bottle 

MA9009  

pH 9.18 buffer, 230 ml bottle 

MA9010  

pH 10.01 buffer, 230 ml bottle 

MA9030  

12.88 mS/cm solution, 230 ml 

MA9015  

Probe storage solution, 230 
ml 

MA9016  

General cleaning solution, 
230 ml  

M10000B  

Rinse solution, 20 ml (25 
pcs.) 

 
 
 
 
 
 
 
 
 
 
WARRANTY

 

This instrument is warranted against defects 
in materials and manufacturing for a period of 
two years from the date of purchase. Probe is 
warranted for 6 months. 
If during this period the repair or replacement 
of parts is required, where the damage is not 
due to negligence or erroneous operation by 
the user, please 
return the parts to either distributor or our 
office and the repair will be effected free of 
charge. 
Damages due to accidents, misuse, 
tampering or lack of prescribed maintenance 
are not covered. 

SPECIFICATIONS 
 
RANGE  

0.00 to 14.00 pH 

 

0.00 to 20.00 mS/cm 

 

0.00 to 10.00 ppt 

 

0.0 to 60.0ºC or 32.0 to 140.0°F 

RESOLUTION  

0.01 pH 

 

0.1 mS/cm 

 

0.01 ppt 

 

0.1°C or 0.1°F 

 

ACCURACY (@25°C)  
 

±0.01pH 

 

±2% FS (EC/TDS) 

 

±0.5°C or ±1°F 

 

TYPICAL EMC DEV.  
 

±0.02 pH 

 

±2% FS (EC/TDS) 

 

±0.5°C or ±1°F 

 

TEMPERATURE 

Automatic from 0 to 

60ºC; 

 

 

COMPENSATION 

ß adj. from 0.0 to 

2.4%/°C 

 

 

pH CALIBRATION 

1 or 2-point 

with automatic buffer recognition 

 

EC CALIBRATION 

Automatic, 1 point 

 

EC/TDS FACTOR (CONV)

Adj. from 

0.45 to 1.00 

 

 

PROBE (included) MA851D/1 

amplified 

pH/EC/TDS/Temperature 

probe with DIN connector and 1 m (3.3’) 
cable. 

 

 

ENVIRONMENT  
 

0 to 50°C (32 to 122°F); 

 

100% RH max. 

 

 

BATTERY TYPE 

1 x 9V alkaline (included) 

BATTERY LIFE 

approx. 300 hours 

 

 

AUTO-OFF 

after 8 minutes 

 

 

DIMENSIONS 

200 x 85 x 50 mm 

 

 

WEIGHT 

260 g (with battery) 

 

 

Отзывы: