GUIDE TO DISPLAY CODES: 
 

This prompt appears for 1 second  each 
time the instrument is turned on.  
 
The dashes “-  -  -” 
indicates 

that the 

meter is in a ready 
state and zeroing can 
be performed.  
 
Sampling In Progress. 
Flashing  “SIP” prompt 
appears each time the 
meter is performing a  
measurement. 
 
“-0.0-”, the meter is in 
a zeroed  state and 
measurement can  be 
performed.  
 
The blinking "BAT" 
indicates 

that the 

battery voltage is 
getting  low and the 
battery  needs to be 
replaced. 
 
“-bA-”, the battery is 
dead and  must be 
replaced. Once this 
indication is displayed, 
the meter will lock up. 
Change the  battery 
and restart the meter. 
 
“Conf”, the meter has 
lost its  configuration. Contact your 
dealer or the nearest Martini  Customer 
Service Center. 
 
 

ERROR MESSAGES 
 
On zero reading 
 

• Blinking “-0.0-” indicates  that the 
zeroing procedure  failed due to a low 
signalto-noise ratio. 
In this case 
press ZERO again. 
• “no L”, the 
instrument can not 
adjust the light 
level. Please check 
that the sample 
does  not contain 
any debris. 
• “L Lo”, there is 
not enough  light to 
perform a 
measurement.  
Please check the 
preparation of the 
zero cuvet. 
• “L Hi”, there is too 
much 

light to 

perform 

measurement. 
Please check the preparation of the zero 
cuvet. 

 
On sample reading 

• “-SA-”, there is too 
much  light for the 
sample 
measurement. 
Please check if  the 
right sample cuvet is 
inserted. 
• “Inv”, the sample 
and the  zero cuvet 
are inverted. 
• “ZErO”, a zero 
reading was 

not 

taken. Follow the instruction  in the 
measurement  procedure for zeroing the 
meter. 
 
• Under range. A blinking  “0.00” 
indicates that the  sample absorbs less 
light  than the zero 
reference. 
Check the procedure 
and  make sure you 
use the same cuvet for 
reference  (zero) and 
measurement. 
• A flashing value of 
the 

maximum 

concentration 
indicates 

an over 

range condition. 
The concentration of 
the sample is beyond 
the 

programmed 

range: dilute  the sample and re-run the 
test. 
 
 
 

BATTERY REPLACEMENT 
 

Battery replacement must only take 
place in a non-hazardous environment. 
Simply rotate the battery cover on the 
back of the meter. 
Detach the 
battery from 
the terminals 
and attach a 
fresh 9V 
battery while 
paying 
attention to 
the 

correct 

polarity. 
Insert the battery and replace the cover. 

ACCESSORIES: 
 
MI505-100  

Ammonia MR Reagent 

 

(100 tests) 

MI0001  

Glass cuvets (2 pcs) 

MI0002  

Caps for cuvets (2 pcs) 

MI0003  

Stoppers for cuvets (2 pcs) 

MI0004  

Tissue for wipping cuvets 

 

(4 pcs) 

MI0005  

9V battery (1 pc) 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

WARRANTY 
 

This instrument is warranted against defects 
in materials and manufacturing for a period of 
two years from the date of purchase. 
If during this period the repair or replacement 
of parts is required, where the damage is not 
due to negligence or erroneous operation by 
the user, please 
return the meter to either distributor or our 
office in the original packing and the repair 
will be free of charge. 
Damages due to accidents, misuse, 
tampering or lack of prescribed maintenance 
are not covered. 