Instruction Manual Mi 190 DO Bench Meter

9

Before proceeding with the calibration, make sure the probe is ready for measurements (see page 7),
i.e. the membrane cap is filled with electrolyte and the probe is connected to the meter and properly
polarized.
For an accurate calibration, it is recommended to wait at least 15 minutes to ensure precise
conditioning of the probe.
Keep the protective cap on during polarization time and remove it for calibration and measurements.
Follow the calibration procedure (see page 10).

SALINITY AND ALTITUDE COMPENSATION

If the sample contains significant concentration of salinity or if you are performing  measure-
ments at an altitude different from sea level, the read out values must be corrected, taking into
account the lower degree of oxygen solubility in these situations (see pages 20-21).
Remember to set the altitude and/or the salinity before taking any D.O. measurements. The
meter will automatically compensate for these factors.

D.O. MEASUREMENTS

Make sure that the instrument has been calibrated and the protective cap has
been removed.
• Immerse the tip of the probe in the sample to be tested. Allow approximately

one minute for the reading to stabilize.

• The Dissolved Oxygen value (in %) is displayed on the primary LCD and the temperature on

the secondary LCD.

2 0

2 0

2 0

2 0

2 0

º C

S a l i n i t y   ( g / l )   a t   S e a   L e v e l

º F

0   g / l

1 0   g / l

2 0   g / l

3 0   g / l

3 5   g / l

0

1 4 . 6 0

1 3 . 6 4

1 2 . 7 4

1 1 . 9 0

1 1 . 5 0

3 2 . 0

2

1 3 . 8 1

1 2 . 9 1

1 2 . 0 7

1 1 . 2 9

1 0 . 9 1

3 6 . 5

4

1 3 . 0 9

1 2 . 2 5

1 1 . 4 7

1 0 . 7 3

1 0 . 3 8

3 9 . 2

6

1 2 . 4 4

1 1 . 6 5

1 0 . 9 1

1 0 . 2 2

9 . 8 9

4 2 . 8

8

1 1 . 8 3

1 1 . 0 9

1 0 . 4 0

9 . 7 5

9 . 4 4

4 6 . 4

1 0

1 1 . 2 8

1 0 . 5 8

9 . 9 3

9 . 3 2

9 . 0 3

5 0 . 0

1 2

1 0 . 7 7

1 0 . 1 1

9 . 5 0

8 . 9 2

8 . 6 5

5 3 . 6

1 4

1 0 . 2 9

9 . 6 8

9 . 1 0

8 . 5 5

8 . 3 0

5 7 . 2

1 6

9 . 8 6

9 . 2 8

8 . 7 3

8 . 2 1

7 . 9 7

6 0 . 8

1 8

9 . 4 5

8 . 9 0

8 . 3 9

7 . 9 0

7 . 6 6

6 4 . 4

2 0

9 . 0 8

8 . 5 6

8 . 0 7

7 . 6 0

7 . 3 8

6 8 . 0

2 2

8 . 7 3

8 . 2 3

7 . 7 7

7 . 3 3

7 . 1 2

7 1 . 6

2 4

8 . 4 0

7 . 9 3

7 . 4 9

7 . 0 7

6 . 8 7

7 5 . 2

2 5

8 . 2 4

7 . 7 9

7 . 3 6

6 . 9 5

6 . 7 5

7 7 . 0

2 6

8 . 0 9

7 . 6 5

7 . 2 3

6 . 8 3

6 . 6 4

7 8 . 8

2 8

7 . 8 1

7 . 3 8

6 . 9 8

6 . 6 1

6 . 4 2

8 2 . 4

3 0

7 . 5 4

7 . 1 4

6 . 7 5

6 . 3 9

6 . 2 2

8 6 . 0

3 2

7 . 2 9

6 . 9 0

6 . 5 4

6 . 1 9

6 . 0 3

8 9 . 6

3 4

7 . 0 5

6 . 6 8

6 . 3 3

6 . 0 1

5 . 8 5

9 3 . 2

3 6

6 . 8 2

6 . 4 7

6 . 1 4

5 . 8 3

5 . 6 8

9 6 . 8

3 8

6 . 6 1

6 . 2 8

5 . 9 6

5 . 6 6

5 . 5 1

1 0 0 . 4

4 0

6 . 4 1

6 . 0 9

5 . 7 9

5 . 5 0

5 . 3 6

1 0 4 . 0

4 2

6 . 2 2

5 . 9 3

5 . 6 3

5 . 3 5

5 . 2 2

1 0 7 . 6

4 4

6 . 0 4

5 . 7 7

5 . 4 8

5 . 2 1

5 . 0 9

1 1 1 . 2

4 6

5 . 8 7

5 . 6 1

5 . 3 3

5 . 0 7

4 . 9 7

1 1 4 . 8

4 8

5 . 7 0

5 . 4 7

5 . 2 0

4 . 9 5

4 . 8 5

1 1 8 . 4

5 0

5 . 5 4

5 . 3 3

5 . 0 7

4 . 8 3

4 . 7 5

1 2 2 . 0

The salinity affects the D.O. concentration, decreasing its value. The table below shows the
maximum oxygen solubility at various temperatures and salinity levels.

Note

Note

Note

Note

Note

:

The relationship between salinity and chlorinity for sea water is given by the equation

below:

Salinity (g/l) = 1.80655 Chlorinity (g/l)

www.milwaukeeinst.com

Содержание Mi 190

Страница 1: ...11 6726 Szeged Hungary Tel 36 62 428 050 Fax 36 62 428 051 e mail sales milwaukeeinst com Milwaukee Instruments Inc 2950 Business Park Drive Rocky Mount NC 27804 USA Tel 1 252 443 3630 Fax 1 252 443...

Страница 2: ...NCTIONALDESCRIPTION B A C D E F G H I 27 WARRANTY WARRANTY WARRANTY WARRANTY WARRANTY This instrument is warranted against defects in materials and manufacturing for a period of 3 years from the date...

Страница 3: ...12Vdc 13 14 2 6 2 6 2 6 2 6 2 6 ACCESSORIES MA 9070 Zero Oxygen Solution 230 ml MA 9071 Refilling Electrolyte Solution 30 ml MA 9311 115VAC to 12VDC converter MA 9310 230VAC to 12VDC converter MA 840...

Страница 4: ......

Страница 5: ...CAN is sent when the instrument is logging Err6 Err8 is sent when the command is incorrect or the instrument is not in measurement mode COMMANDS REQUIRING AN ANSWER The instrument will answer for the...

Страница 6: ......

Страница 7: ...e temperature recalibration should be performed For an accurate recalibration contact your dealer or the nearest Milwaukee Customer Service Center or follow the instructions bellow Prepare a vessel co...

Страница 8: ...11 7 11 2 10 8 10 4 10 0 9 7 9 3 9 0 8 9 32 0 2 13 8 13 3 12 8 12 4 11 9 11 5 11 0 10 6 10 2 9 9 9 5 9 2 8 8 8 5 8 4 35 6 4 13 1 12 6 12 2 11 7 11 3 10 9 10 5 10 1 9 7 9 3 9 0 8 7 8 4 8 0 7 9 39 2 6...

Страница 9: ...S e a L e v e l F 0 g l 1 0 g l 2 0 g l 3 0 g l 3 5 g l 0 1 4 6 0 1 3 6 4 1 2 7 4 1 1 9 0 1 1 5 0 3 2 0 2 1 3 8 1 1 2 9 1 1 2 0 7 1 1 2 9 1 0 9 1 3 6 5 4 1 3 0 9 1 2 2 5 1 1 4 7 1 0 7 3 1 0 3 8 3 9 2...

Страница 10: ...e was stored and the temperature of the sample the longer the time will be Note If is displayed the D O probe is not properly connected or the temperature is out of range This also indicates the posib...

Страница 11: ...witch the meter on by pressing the ON OFF switch For an accurate calibration it is recommended to wait for at least 15 minutes to ensure precise conditioning of the probe Remove the protective cap fro...

Страница 12: ...ty factor The altitude factor The dEL message on the primary LCD and the record number on the secondary LCD along with ACCEPT tag blinking Notes Notes Notes Notes Notes If one of the arrow keys is pre...

Страница 13: ...tores the value and adjusts the slope point The instrument stores the slope calibration data and returns to measurement mode Note Note Note Note Note If the reading is not close to the selected value...

Страница 14: ...measurement mode e g 18 records free If there are less than 6 memory locations remaining the record number and Lo message will blink to alert the user 15 If the log space is full FULL LOC message wil...

Отзывы: