Instruction Manual Mi 190 DO Bench Meter

11

D.O. CALIBRATION

Calibrate the instrument frequently, especially if high accuracy is required.
The instrument can be calibrated in maximum 2 points: 0.0% (zero calibration

zero calibration

zero calibration

zero calibration

zero calibration) and

100.0% (slope calibration

slope calibration

slope calibration

slope calibration

slope calibration).

The zero calibration of the Mi 190

Mi 190

Mi 190

Mi 190

Mi 190 is very stable, therefore this procedure needs to be

performed only whenever the probe or the membrane is replaced.
However, because the slope calibration is more critical, it is recommended to perform this
procedure every week.

INITIAL PREPARATION

• Pour small quantities of MA 9070  Zero Oxygen solution into a

beaker. If possible, use a plastic beaker to minimize any EMC interfer-
ences.

• Make sure the probe is ready for measurements (see probe prepara-

tion on page 7), i.e. the membrane is filled with electrolyte and the
probe is connected to the meter.

• Switch the meter on by pressing the ON/OFF switch.

• For an accurate calibration, it is recommended to wait for at least

15 minutes to ensure precise conditioning of the probe.

• Remove the protective cap from the D.O. probe.

• Set the appropriate altitude factor (see page 21). Make sure the

salinity factor is set to zero (see page 20).

ZERO CALIBRATION

• Submerge the probe into MA 9070     zero oxygen solution and stir

gently for 2-3 minutes.

MI9070

MI9070

1 8

1 8

1 8

1 8

1 8

SETUP

Setup mode allows viewing and modifying the following parameters:

• Salinity Compensation

• Altitude Compensation

• Current Time (hh:mm)

• Current Date (MM.DD.YYYY)

• Beep Status

• Baud Rate (serial communication)

• Instrument ID

• Temperature Unit
To enter SETUP mode, press and hold the SETUP key for about 1 second while in normal
measurement mode.

Select the desired setup parameter using the UP and DOWN arrow keys.

Press the CAL key if you want to change the item value. The selected item (e.g. hour, in setting
up the correct time) and “ACCEPT” tag will start blinking.

Press the arrow     keys to change the displayed value.

If there is another item to be set (e.g. minutes), press the RANGE key. The other item will start
blinking.

www.milwaukeeinst.com

Содержание Mi 190

Страница 1: ...11 6726 Szeged Hungary Tel 36 62 428 050 Fax 36 62 428 051 e mail sales milwaukeeinst com Milwaukee Instruments Inc 2950 Business Park Drive Rocky Mount NC 27804 USA Tel 1 252 443 3630 Fax 1 252 443...

Страница 2: ...NCTIONALDESCRIPTION B A C D E F G H I 27 WARRANTY WARRANTY WARRANTY WARRANTY WARRANTY This instrument is warranted against defects in materials and manufacturing for a period of 3 years from the date...

Страница 3: ...12Vdc 13 14 2 6 2 6 2 6 2 6 2 6 ACCESSORIES MA 9070 Zero Oxygen Solution 230 ml MA 9071 Refilling Electrolyte Solution 30 ml MA 9311 115VAC to 12VDC converter MA 9310 230VAC to 12VDC converter MA 840...

Страница 4: ......

Страница 5: ...CAN is sent when the instrument is logging Err6 Err8 is sent when the command is incorrect or the instrument is not in measurement mode COMMANDS REQUIRING AN ANSWER The instrument will answer for the...

Страница 6: ......

Страница 7: ...e temperature recalibration should be performed For an accurate recalibration contact your dealer or the nearest Milwaukee Customer Service Center or follow the instructions bellow Prepare a vessel co...

Страница 8: ...11 7 11 2 10 8 10 4 10 0 9 7 9 3 9 0 8 9 32 0 2 13 8 13 3 12 8 12 4 11 9 11 5 11 0 10 6 10 2 9 9 9 5 9 2 8 8 8 5 8 4 35 6 4 13 1 12 6 12 2 11 7 11 3 10 9 10 5 10 1 9 7 9 3 9 0 8 7 8 4 8 0 7 9 39 2 6...

Страница 9: ...S e a L e v e l F 0 g l 1 0 g l 2 0 g l 3 0 g l 3 5 g l 0 1 4 6 0 1 3 6 4 1 2 7 4 1 1 9 0 1 1 5 0 3 2 0 2 1 3 8 1 1 2 9 1 1 2 0 7 1 1 2 9 1 0 9 1 3 6 5 4 1 3 0 9 1 2 2 5 1 1 4 7 1 0 7 3 1 0 3 8 3 9 2...

Страница 10: ...e was stored and the temperature of the sample the longer the time will be Note If is displayed the D O probe is not properly connected or the temperature is out of range This also indicates the posib...

Страница 11: ...witch the meter on by pressing the ON OFF switch For an accurate calibration it is recommended to wait for at least 15 minutes to ensure precise conditioning of the probe Remove the protective cap fro...

Страница 12: ...ty factor The altitude factor The dEL message on the primary LCD and the record number on the secondary LCD along with ACCEPT tag blinking Notes Notes Notes Notes Notes If one of the arrow keys is pre...

Страница 13: ...tores the value and adjusts the slope point The instrument stores the slope calibration data and returns to measurement mode Note Note Note Note Note If the reading is not close to the selected value...

Страница 14: ...measurement mode e g 18 records free If there are less than 6 memory locations remaining the record number and Lo message will blink to alert the user 15 If the log space is full FULL LOC message wil...

Отзывы: