“ON/OFF” button: the LCD will show the 
“OK 1” message for 1 second and then 
returns to the normal mode. If using pH 
4.01 or 10.01 (or 9.18) buffer for the first 
calibration point, the meter will exit the 
calibration mode after recognizing the 
buffer value and showing the “OK 1” 
message for 1 second. 
•   To quit the calibration procedure and 
return to the last data, press “ON/OFF” 
after entering the calibration mode and 
before the first point is accepted. The 
secondary LCD shows the “ESC” 
message for 1 second and the meter 
returns to the normal mode. 
•   To clear a previous calibration and 
reset to the default value, press the 
“SET” key after entering the calibration 
mode and before the first point is 
accepted. The    secondary LCD shows 
the “CLR” message for 1 second, the 
meter resets to the default calibration 
and the “CAL” tag on the LCD 
disappears. 
 
 
 

TAKING ORP MEASUREMENTS 
 

• 

Remove the protective cap and 

immerse the electrode in 

MA9016 

cleaning solution for a few minutes. 
•   Select the ORP (mV) mode with the 
“SET” key. 
•   Place the probe into the sample to be 
tested, stir gently and wait for reading 
stabilization (the clock symbol on the 
LCD disappears). 

Note

: The ORP range is factory 

calibrated. Contact your dealer for 
recalibration, if needed. 

Note

: ATC tag turns off because the mV 

readings are not temperature 
compensated. 

 

BATTERY REPLACEMENT 
 

The meter shows the remaining battery 
percentage every time it is switched on. 
When the battery level is below 5%, the 
battery symbol on the bottom left of the 
LCD lights up to indicate a low battery 
condition. 
The meter is also provided with BEPS 
(Battery Error Prevention System), which 
automatically switches the meter off if the 
battery level is so low to cause 
erroneous readings. It is recommended 
to replace immediately the battery.  
Turn the meter off, 
remove  the battery 
compartment cover 
from the rear of the 
meter and replace the 
rundown 9V battery 
with a new one. 
Install the battery 
while paying attention 
to its polarity and 
reattach the cover. 
 

WARRANTY 

This instrument is warranted against defects in 
materials and manufacturing for a period of two 
years from the date of purchase. Probe is 
warranted for 6 months. 
If during this period the repair or replacement of 
parts is required, where the damage is not due to 
negligence or erroneous operation by the user, 
please return the parts to either distributor or our 
office and the repair will be effected free of 
charge. 
Damages due to accidents, misuse, tampering or 
lack of prescribed maintenance are not covered. 
 

 
 
 
 
 
 
 
 

ACCESSORIES 
 
MA923D/1 

Combination amplified 
pH/ORP/Temperature 
probe with DIN 
connector and 1 m cable 

M10004B  

pH4.01 buffer solution, 
20 ml sachet, 25 pcs. 

M10006B  

pH6.86 buffer solution, 
20 ml sachet, 25 pcs. 

M10007B  

pH7.01 buffer solution, 
20 ml sachet, 25 pcs. 

M10009B  

pH9.18 buffer solution, 
20 ml sachet, 25 pcs. 

M10010B  

pH10.01 buffer solution, 
20 ml sachet, 25 pcs. 

MA9004  

pH4.01 buffer solution, 
230 ml bottle 

MA9006  

pH6.86 buffer solution, 
230 ml bottle 

MA9007  

pH7.01 buffer solution, 
230 ml bottle 

MA9009  

pH9.18 buffer solution, 
230 ml bottle 

MA9010  

pH10.01 buffer solution, 
230 ml bottle 

MA9015  

Electrode storage 
solution, 230 ml bottle 

MA9016  

Electrode cleaning 
solution, 230 ml bottle 

M10000B  

Electrode rinse solution, 
20 ml sachet, 25 pcs. 

 

 
 

 
 
 
 
 
 
 

SPECIFICATIONS 

 
 

 

RANGE  

-2.00 to 16.00 pH 

 

-2000 to + 2000 mV 

 

-5.0 to 105.0 ºC  

 

or 23.0 to 221.0 °F 

RESOLUTION  

0.01 pH / 1 mV 

 

0.1 °C or 0.1 °F 

ACCURACY 
 (@25°C)  

±0.02 pH / ±2 mV 
±0.5°C up to 60°C; 
±1°C outside 

±1°F up to 140 °F; ±2°F 
outside 

TYPICAL EMC  
DEVIATION  

±0.02 pH / ±2 mV 

 

±0.2 °C or ±0.4 °F 

TEMPERATURE  
COMPENSATION 

Automatic, from -5 to 
80°C 

pH CALIBRATION 

Automatic, 1 or 2- 

 

point 

ORP CALIBRATION 

Factory  

 

calibrated 

PROBE (included) MA923D/1 

amplified 
pH/ORP/temperature 
probe 

ENVIRONMENT  

0 to 50°C; 100% RH 
max. 

BATTERY  

1 x 9V alkaline 
(included) 

BATTERY LIFE  

approx. 500 hours of 
use 

AUTO-OFF  

after 8 minutes of 
non-use 

DIMENSIONS  

200×85×50 mm 

WEIGHT  

260 g (with battery) 

 
(*) 

The temperature range is limited to 80°C 

(176°F) if using the 

MA923D/1 

probe.

 