PRELIMINARY EXAMINATION

Thank you for choosing Milwaukee Instruments!
While taking out the meter from the packaging, make sure that it was 

not damaged during shipping. Please notify your closest Milwaukee 

office if it was damaged.
The EC40 is delivered in a plastic tube and is supplied complete with:

 

y

3 x 1.5V AA batteries

 

y

Instrument quality certificate

 

y

Instruction manual

Note: Keep all packing material until you have verified that the 

instrument works correctly. Defective or damaged items must be 

returned in their original packing material.

GENERAL DESCRIPTION

The EC40 Nutrient stick is designed to measure the EC (conductivity) of 

the nutrient solutions used in agriculture and hydroponic applications. 

It features a visual alarm that blinks if the measured value is above the 

set point. The EC40 is built in a compact, waterproof and floating casing.

FEATURES

 

y

Readings are displayed on an easy to read graph bar with 20 LEDs.

 

y

The readings are displayed in four measuring scales: CF, EC, PPM500 

and PPM700.

 

y

When the reading is out of range, the top or bottom LED blinks.

 

y

The alarm set point can be easily adjusted.

 

y

The meter is factory calibrated and does not require any calibration.

 

y

The meter switches on and off automatically.

 

y

Battery level indicator activates when you insert the batteries into 

the instrument.

 

y

Its compact, waterproof and floating design makes the meter water 

resistant and ideal to mix the nutrient solution in the tank.

 

y

The probe is protected by a crown-shaped ring which is detachable 

so it is easy to clean.

 

y

The meter is supplied with a wrist strap.

 

y

Essential operational instructions are printed on the meter.

 

y

The meter has a graphite EC probe with a built-in temperature sensor, 

housed in a stainless-steel casing, which compensates temperature 

in seconds.

SPECIFICATIONS

Range

0.2 to 4 mS/cm
2 to 40 CF
140 to 2800 ppm 0.7
100 to 2000 ppm 0.5

Resolution

0.1 mS/cm
1 CF
70 ppm (factor: 700 ppm = 1 mS)
50 ppm (factor: 500 ppm = 1 mS)

Accuracy

± 4% of reading ± 1 resolution point

Calibration 

Factory calibrated

Probe

Graphite electrode in ABC + PC body

Temperature 

compensation

Automatic from 5.0 to 50.0 ºC  

(41.0 to 122.0 ºF) 

Battery type

3 x 1.5V AA alkaline

Battery life

about 3 years (at 10 measurements per day) 

Measurement display

20 blue LEDs bar

Environment

0 to 50 ºC (32.0 to 122.0 ºF); max. RH 95%

Dimensions

420 mm (17”) length, 30 mm (1.2”) diameter

Weight with batteries

255 g. (9 oz.)

Casing protection

IP67, floating

OPERATIONAL GUIDE

When immersed, the meter starts measuring automatically and 

displays current Electrical Conductivity level. Auto shut-off time is set at 

20 seconds after the measurement has stabilized.

Switching the Meter On

The meter switches on when the batteries are placed inside. 

It performs a self-test to identify possible LED issues. The LEDs turn 

on one by one from the bottom to the top and then reverse top to 

bottom. Then, the LED bar shows the voltage level of the batteries for 

a few seconds. If the top LED is on it means that batteries are fully 

charged. If the bottom LED is on, it indicates low charge. In this case, 

please change the batteries. 

Measurement

1.  Place the head of the probe in the nutrient solution. The meter 

switches on automatically and starts measuring.

2.  An LED light turns on once the reading has stabilized. The readings 

are displayed in four measuring scales: CF, EC, PPM500 and PPM700. 

If 2 LEDs are on, the measured value is between those two values.

3.  Continuously blinking LED/LEDs indicate that the measured value is 

over the alarm set point.

4.  The meter remains on for approximately 20 seconds. If the EC value 

changes, auto shut-off time is automatically reset. The meter stops 

measuring if the read value drops below minimum value.

5.  The top/bottom LED is blinking when the measurement is over/

under range.

Alarm Set Up

1.  Rotate the battery cap counter-clockwise and remove it.

2.  Push the SETUP button once.
3.  The LED turns on and displays the last set alarm value.
4.  To change the alarm value, push and hold the SETUP button. The 

LED bar increases gradually. Release the button to save the new 

alarm value.

5.  To disable the alarm, release the SETUP button when the top and 

bottom LEDs are on.

6.  Put back the battery cap.

Calibration

The meter is factory calibrated and does not require user calibration.

CLEANING & MAINTENANCE

Frequent cleanings are recommended to ensure accurate 

measurements.

1.  Twist the crown-shaped ring 30 degrees counter-clockwise then 

remove it.

2.  Clean the head of the probe under running tap water.
3.  If required, clean the graphite sensors with a couple drops of dish 

liquid detergent, and scrub off any residue with a toothbrush.

4.  Rinse the probe under running tap water.

STORAGE

Make sure that the probe is clean and dry before storing it in its 

protective packaging.