14
Table 5 Capacitance Test:
Step
Source
Reading
1
3.6uF
3.3uF to 3.9uF
2
80.0uF
78.3uF to 81.7uF
3
320uF
312uF to 328uF
4
1000uF
979uF to 1021uF
Testing the Temperature Function
To verify the accuracy of the temperature function, do the following:
1. Connect the Calibrator to
V
Ω
Hz
and
COM
on the Meter.
2. Wait for at least 30 minutes.
2. Turn the rotary switch to Temperature.
3. Apply the inputs for step 1-11 in Table 6.
4. Compare the Meter display readings to the display readings in Table 6.
5. If the display reading falls outside of the range shown in Table 6, the Meter does not meet
specification.
Table 6 Temperature Test:
Step Input
Mode
Reading
1
0.0 degree
Celsius
-2.0 to 2.0
2
-10.0 degree
Celsius
-12.1 to -7.9
3
-40.0 degree
Celsius
-42.4 to -37.6
4
100.0 degree
Celsius
97.0 to 103.0
5
400.0 degree
Celsius
394.0 to 406.0
6
0.0 degree
Fahrenheit
-3.0 to 3.0
7
-10.0 degree
Fahrenheit
-6.9 to -13.1
8
-40.0 degree
Fahrenheit
-36.6 to -43.4
9
32.0 degree
Fahrenheit
28.7 to 35.3
10
400.0 degree
Fahrenheit
393.0 to 407.0
11
750.0 degree
Fahrenheit
739.5 to 760.5