Usage Guidelines

Always use the Millex

®

-FG

50

 vent filter and silicone tubing included

with the pump when applying vacuum. This will prevent liquids or
vapors from entering the Millivac Maxi Pump. A properly attached
Millex-FG

50

 vent filter is illustrated in Figure 1.

For maximum protection of the pump, use a Millex-FG

50

 vent filter

AND a vacuum trap to protect the pump from vapors and gases.
The vacuum trap consists of a 1 liter filtering flask (Millipore cat.
no. XX10 047 05), No. 8 stopper (XX10 047 08) and silicone tubing.
Additional tubing is available as Millipore cat. no. XX80 000 24.
This assembly is illustrated in Figure 2.

Never pump or draw liquids through the pump. This will damage
the pump. The filter flask and vacuum trap should be emptied after
each use.

Figure 1. Proper use of Millex-FG

50

 filter.

Millex-FG

50

filter unit

Filter

flask

Vacuum

pump

Filter

holder

Figure 2. Proper use of vacuum trap and Millex-FG

50

 filter.

Filter

holder

Filter

flask

Vacuum

trap

Vacuum

pump

Millex-FG

50

filter unit

Содержание Millivac Maxi SD1P014M04

Страница 1: ...Millivac Maxi Diaphragm Pump Catalogue No SD1P014M04 230 V 50 Hz ...

Страница 2: ...y appear in this document This manual is believed to be complete and accurate at the time of publication In no event shall Millipore Corporation be liable for incidental or consequential damages in connection with or arising from the use of this manual 2005 MILLIPORE CORPORATION P90725 Rev 06 05 Millipore and MIllex are registered trademarks of Millipore Corporation Millivac is a trademark of Mill...

Страница 3: ...ng oil vapors that might pollute the transferred evacuated or compressed air The design of this small pump includes a patented diaphragm for maintenance free service for a minimum of one year The design is also gastight with a leakage rate of approximately 6 10 3 mbar l s CAUTION Always use a hydrophobic vent filter or a vacuum flask water trap in conjunction with the pump See Figures 1 and 2 Neve...

Страница 4: ...es The vacuum trap consists of a 1 liter filtering flask Millipore cat no XX10 047 05 No 8 stopper XX10 047 08 and silicone tubing Additional tubing is available as Millipore cat no XX80 000 24 This assembly is illustrated in Figure 2 Never pump or draw liquids through the pump This will damage the pump The filter flask and vacuum trap should be emptied after each use Figure 1 Proper use of Millex...

Страница 5: ...lace Do not abuse the cord Replace the Millex FG50 vent filter SLFG 050 10 if necessary Inspect hose plug and cord for signs of damage before use Do not use if a deficiency is found Never operate a damaged unit Contact Millipore for replacement parts see Technical Assistance section for details This vacuum pump needs no lubrication Applying oil to any part could result in polluted air delivery to ...

Страница 6: ...low to identify the parts of the Millivac Maxi pump Pump Components A Pump heads B Pump head connector C Digital vacuum gauge D Digital vacuum gauge connector E Nut F Vacuum port G Bleeder valve H Regulator knob I On off switch J Tubing K Millex FG50 filter unit L Electrical pigtail A B D E F G H I C J K L ...

Страница 7: ...n slide the nut onto the threads of the pump head connector and hand tighten by turning the nut clockwise 4 Tighten nut securely using a 3 4 or 19 mm wrench Attach the Tubing 1 Attach the supplied tubing to the pump by pressing it onto the vacuum port on the digital vacuum gauge 2 Insert the Millex FG50 filter device into other end of the tubing 3 Connect tubing from appropriate filtration device ...

Страница 8: ...e Press the Select key to turn the digital vacuum gauge on When first turned on the gauge will flash the full scale pressure range and the software version year week After that the actual pressure displays at the top of the screen and the last measured MAX value shows at the bottom of the screen View the MIN value for 5 seconds by pressing the Enter key NOTE The digital vacuum gauge will turn on a...

Страница 9: ...essure OFF Press the Select key twice to choose Off Press Enter to turn the digital vacuum gauge off MANO Press the Select key three times and then press Enter to choose Mano Mano allows access to a number of functions Press Select until the de sired function is displayed and then press Enter to select the function Zero Set Sets a new zero reference at atmospheric pressure Use for measurement of r...

Страница 10: ...gauge turn the regulator knob counter clockwise to the open position Maintenance Under normal operating conditions and using proper handling proce dures the Millivac Maxi Pump should provide many hours of trouble free operation These durable units are built for quiet continuous duty operation Millivac Maxi Pumps are 100 oil free The pump employs a non lube piston and cylinder No maintenance is nec...

Страница 11: ...mm Weight 3 95 kg Voltage 50 Hz AC power Maximum speed 50 rpm Operating pressure 0 5 bar Flow rate 16 L min Ultimate vacuum 20 mbar abs Operating temperature 5 C to 40 C Motor protection IP20 Connectors for tube ID 6 mm CE Declaration of Conformity Millipore Corporation certifies that this product complies with the fol lowing European Union Directives Electromagnetic Compatibility EMC Directive 89...

Страница 12: ...ns for a period of one year from shipment of the products MILLIPORE MAKES NO OTHER WARRANTY EXPRESSED OR IMPLIED THERE IS NO WARRANTY OF MERCHANT ABILITY OR FITNESS FOR A PARTICULAR PURPOSE The warranty provided herein and the data specifications and descriptions of Millipore products appearing in Millipore s published catalogues and product literature may not be altered except by express written ...

Отзывы: