Page 17 of 32
Note: the switch starts blinking for 15
minutes while the following operations
occur:
1. The plunger is partially pulled down and
the sanitizer fills the pump internal flow
path.
2. About 10 mL of sanitizer remains in the
syringe
3. The sanitizer remains in contact with
the internal flow for 15 minutes
4. The plunger is pulled down completely
5. The water in the funnel is pulled to
rinse the pump
6. The switch stops blinking
7. The pump switches off if not in use for
thirty minutes
• When the light is steady or if the pump is
switched off, repeat the operation for the
second flow path
• Discard the sanitization kit consumables
• Remove the sanitization kit and clean it
using a wipe moistened with one of the
surface sanitizers listed in the cleaning &
maintenance section
Note: the sanitization kit can be
autoclaved at 121 °C for 15 min or
at 134 °C for 5 min.
4.8 Filtration head
complete cleaning
Materials:
1. Pump head
2. Gasket removal tool (to be found in the
pump spare part kit or together with
the sanitization kit)
3. Sanitizer (see table)
• Remove the filtration head from the
pump
• Remove the clear external silicone gasket
at the exterior of the head
Figure 36: external gasket removal
Optional: do the same with the external
colored ring (if used)
Figure 37: colored ring removal