T
ROUBLESHOOTING
- 40 -
PROBLEM
POSSIBLE CAUSES
WHAT TO DO
The system may have
been in STANDBY
mode for a while.
The Product resistivity is
below the Resistivity
Setpoint set in menu
option C02.
Dispense water for few
minutes
Replace the SimpliPak
Cartridge. See
Section 6-2
How to Replace the SimpliPak
Cartridge.
UV System only
The UV Lamp Timer is
exhausted.
Replace the UV Lamp. See
Section 6-5 How to Replace
the UV Lamp.
After a new UV Lamp has
been installed, reset the UV
Timer. See
Section 6-6 How to
View or Reset the UV Lamp
Timer (C03).
UV System only
The UV Lamp is not
installed correctly or not
installed at all.
Power OFF the system and
reinstall the UV Lamp.
If the red UV Lamp Alarm is
still displayed, then contact
Millipore.