![Millipore ICW-3000 Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/millipore/icw-3000/icw-3000_user-manual_1796917012.webp)
P
RE
I
NSTALLATION
- 6 -
Chapter 3
P
RE
I
NSTALLATION
3-1
I
NSTALLATION
R
EQUIREMENTS
P
RETREATED WATER
R
EQUIREMENTS
Type of pretreated water
Only Elix or distilled water*
Temperature
5
C < T < 35
C
E
NVIRONMENTAL
R
EQUIREMENTS
Indoor Use Only
Ambient Storage Temperature
5
C < T < 40
C
Ambient Operating Temperature
5
C < T < 40
C
Relative Humidity
Should not exceed 80% for temperatures below 31
C.
Should stay within 50% to 80% between 31
C and 40
C.
Altitude <
3000
metres
Installation Category
II
Pollution Degree
2
*Note: Product water quality specifications may not be met if the ICW-3000 System is fed with DI water,
and it could shorten the lifetime of your consumables.
3-2
U
NPACKING THE
ICW-3000
S
YSTEM
W
HAT
’
S
I
NSIDE
?
Open the ICW-3000 System Shipping Box. Use the checklist included in the Accessories Bag to make sure
all items were shipped and are accounted for. It is highly suggested to become familiar with the items that
are shipped since these will be used in the Installation section of this manual.
Contact Millipore if an item is missing.
Содержание ICW-3000