Millipore Amicon Ultra-2 Скачать руководство пользователя страница 2

Amicon

®

 Ultra-2 Centrifugal Filter Devices

2

Required Equipment

Centrifuge with swinging bucket or fixed angle rotor with wells/carriers that can accommodate 17 mm × 100 mm tubes (same 

well/carrier size as for Amicon

®

 Ultra-4 devices and the former Centricon

®

 device).

CAUTION: 

 To avoid damage to the device during centrifugation, make sure it is properly assembled and seated at the bottom of 

the rotor. The rim of the concentrate collection tube should be inside the rotor well. Check clearance before spinning.

Suitability

Preliminary recovery and retention studies are suggested to ensure suitability for intended use. See the “How to Quantify 

 

Recoveries” section.

Device Storage

Store at room temperature.

Prerinsing

The ultrafiltration membranes in Amicon

®

 Ultra-2 devices contain trace amounts of glycerine. If this material interferes with 

 

analysis, pre-rinse the device with buffer or Milli-Q

®

 water. If interference continues, rinse with 0.1 N NaOH followed by a second 

spin of buffer or Milli-Q

®

 water.

CAUTION:

  Do not allow the membrane in Amicon

®

 Ultra filter devices to dry out once wet. If you are not using the device 

 

immediately after pre-rinsing, leave fluid on the membrane until the device is used.

How to Use Amicon

®

 Ultra-2 Centrifugal Filter Devices

1.  Insert the Amicon

®

 Ultra-2 device into the filtrate collection tube, making sure that the device is fully seated in the tube.

2. 

Add up to 2 mL of sample to the device and cover with concentrate collection tube. Push the tube firmly onto the device.

WARNING: 

 Failure to fully seat the device in the filtrate collection tube and push the concentrate collection tube firmly onto 

the device may result in the device breaking during centrifugation. See figure below.

10K

4

3

2

00

00

00

10K

4

3

2

1

00

20

00

00

00

Make sure both tubes are 

fully seated

 onto device

Add sample

3

2

1

00 

20

00

00

00

10K 

3. 

Place filter device into the centrifuge rotor with one membrane panel facing the center of the rotor (one panel facing up and the 

other panel facing down). Make sure the device is seated on the bottom of the rotor well and that the rim of the concentrate 

collection tube is completely inside the well. See figures below. Counterbalance with a similar device.

10K

4

3

2

1

00

20

00

00

00

Spin to 

 

concentrate

10K

 

00 

00 

White membrane 

panel facing  

center of rotor

Orient device correctly in rotor

Seat device fully in well

Rim of tube 

Concentrate tube 

should be

 

inside

 

4. 

Spin for approximately 10–60 minutes depending on the MWCO of the device used: 

4,000 × g maximum 

when using a swinging bucket rotor

7,500 × g maximum 

when using a fixed angle rotor

NOTE:    

When spinning viscous solutions such as undiluted serum or plasma, do not exceed 5,400 x g.

Refer to Figures 1 and 2 and Tables 2 and 3 for typical spin times.

5. 

Remove the assembled device from the centrifuge and separate the Amicon

®

 Ultra filter device from the filtrate collection tube.

6.  To recover the concentrated solute, invert the Amicon

®

 Ultra filter device and concentrate collection tube. Place in centrifuge 

and counterbalance with a similar device. Spin for 2 minutes at 1,000 × g to transfer the concentrated sample from the device 

to the tube.

NOTE:

 For optimal recovery, perform the reverse spin immediately.

Содержание Amicon Ultra-2

Страница 1: ...pun in a swinging bucket or fixed angle rotor Amicon Ultra 2 devices are supplied non sterile and are for single use only The Amicon Ultra 2 product line includes 5 different cutoffs Molecular Weight...

Страница 2: ...Failure to fully seat the device in the filtrate collection tube and push the concentrate collection tube firmly onto the device may result in the device breaking during centrifugation See figure bel...

Страница 3: ...mL protein in 100 mM NaCl 50 L of 40 mg mL protein in 100 mM NaCl 100 mM NaCl 50 L of 40 mg mL protein in 12 3 mM NaCl 12 3 mM NaCl Add 1 95 mL of 10 mM NaCl or exchange buffer 2 mL of 1 mg mL protei...

Страница 4: ...10 to 20 minutes of centrifugation the lowest concentrate volume 30 70 L is reached after spinning for 10 to 60 minutes Figure 1 Typical Filtrate Volume vs Spin Time for Amicon Ultra 2 Device Swingin...

Страница 5: ...nd 10K BSA for 30K and 50K and IgG for 100K n 8 Shaded volumes were used for the calculation of protein recovery in Table 5 Table 3 Typical Concentrate Volume Concentration Factor v Spin Time Fixed An...

Страница 6: ...nging bucket rotor 4 000 g or 35 fixed angle rotor 7 500 g 2 mL starting volume room temperature n 12 Factors that determine sample recovery include the nature of the protein solute relative to the de...

Страница 7: ...te collection tube and concentrate collection tube before use 2 Fill filter device with solution and reweigh 3 Assemble device in filtrate collection tube and centrifuge per instructions 4 Collect the...

Страница 8: ...rum or plasma do not exceed 5 400 x g Active membrane area 1 cm2 Hold up volume 5 L Dimensions Filter device and tube Length concentration mode device in filtrate tube 119 7 mm 4 71 in Length recovery...

Страница 9: ...ent 0 1 Tergazyme detergent 1 Lubrol PX detergent 0 1 Triton X 100 surfactant 0 1 Nonidet P 40 surfactant 2 Tween 20 surfactant 0 1 Sodium deoxycholate 5 Organic solvents Acetone Not recommended Ethyl...

Страница 10: ...FC900396 10K 8 24 96 500 UFC501008 UFC501024 UFC501096 UFC5010BK UFC201024 UFC801008 UFC801024 UFC801096 UFC901008 UFC901024 UFC901096 10K IVD 8 24 96 UFC801008D UFC801024D UFC801096D UFC901008D UFC90...

Страница 11: ...office nearest you go to www sigmaaldrich com offices Technical Assistance Visit the tech service page on our web site at www sigmaaldrich com techservice Standard Warranty The applicable warranty fo...

Отзывы: