20 of 24

Message

General Cause

Corrective Action

Warning High concentration

Concentration of sample  
is too high

Dilute sample so that it is no higher than recommended maximum 
operating range guideline. See 

Specifications, p .22

.

Lost Start

Sample volume too small, 
sensor not fully immersed 
in solution while sample is 
loading, or air bubble in sensor

Make sure sample volume is ≥ 100 µL. Keep sensor fully 

immersed while screen displays. Keep the Sensor Submerged.

False start

Sensor malfunction

Replace with unused sensor.

False stop

Start/Stop short

Previously used sensor

Sensors are not reuseable, replace with unused sensor.

Electrode short

Sensor malfunction

Replace sensor.

Aperture block

Sample population size  
exceeds aperture size

Dilute to reduce incidences, but may not eliminate. Use 
D-PBS for diluting solution. See 

Cell, Bead, and Particle 

Suspensions, p .12 .

Viscous sample  
(e.g. serum-enriched media)

Keep sensor submerged in sample until Count Complete appears 
on the display. 
Use D-PBS for diluting sample.

Cell Counter sensor is blocked

Sample concentration is too high; dilute cell sample more.
Ensure that the cells are in a single-cell suspension. Break 
clumps by pipetting up and down with a standard pipettor or 
pre-filter using Steriflip

®

 filter unit with 20 µm nylon net.

Air bubble in sensor due to: 
Pre-mature pressing of OK 
button

Fully submerge sensor in sample before pressing OK button.

Air bubble in sample

Mix sample gently to avoid air bubbles.

Sensor lifted from sample 
before screen displays 

Sample 

Loaded

Keep sensor submerged in sample while screen displays Keep 
the Sensor Submerged.

Wrong diluting solution

Refer to list of acceptable diluting solutions in 

Cell, Bead, and 

Particle Suspensions, p .12

.

Debris from diluting solution

Diluents can be filtered to remove debris using ≥0.2 µm using 

Stericup

®

, Steritop

®

 or Steriflip

®

 filters. See 

Product Ordering 

for catalogue numbers. 

Open Start

Issues detected upon sensor 
insertion

Re-insert sensor. Try new sensor, if problem persists, contact 
Tech Service.

Open stop

Previously used sensor 
detected

Re-insert sensor. Try new sensor, if problem persists, contact 
Tech Service.

System stopped 

Error occurred or Menu  
bottom pressed during  
counting to abort test

Replace sensor and retest.

No histograms  
currently stored

No stored data 

The instrument is new or all histograms have been deleted.

Unsuccessful file transfer 
Some histograms  
failed to export

Low battery 
or USB thumb drive  
not inserted 
or USB thumb drive  
not fully inserted  
or USB thumb drive is 
malfunctioning or full 
or USB thumb drive is  
not formatted to FAT32 

Charge battery and re-try file transfer.

Push USB thumb drive into the Scepter™ 3.0  
USB port firmly. 

Check the USB thumb drive to ensure it is  
formatted FAT32, has sufficient space available  
and is functioning properly. 

See 

USB Thumb Drive Formatting .

 

Can not mount  
USB thumb drive

Содержание PHCC30000

Страница 1: ...perates as MilliporeSigma in the U S and Canada The life science business of Merck KGaA Darmstadt Germany operates as MilliporeSigma in the US and Canada SigmaAldrich com User Guide Scepter 3 0 Handhe...

Страница 2: ...r computer easily No more cables The new Charging Station stores Scepter 3 0 out of the way when not in use freeing up valuable hood or lab bench space This system is intended for research use only Co...

Страница 3: ...nstrument to vibrations Vibrations may cause instrument malfunction or damage Do not autoclave or expose to high temperature To avoid damage to internal electronic components do not spray instrument w...

Страница 4: ...s the sensor Tech Service Port For Service use only USB Port Connects instrument to USB thumb drive Ergonomic Hook Allows comfortable handling of the instrument Secures the instrument to charger Contr...

Страница 5: ...the Recharging screen when placed in the Charging Station the following steps must be followed to start charging the Scepter 3 0 battery 1 Place Scepter 3 0 into the Charging Station 2 While on the s...

Страница 6: ...kable surface 1 Use double sided mounting tape 2 Remove tape liner from one side of tape and affix to back of mounting plate as shown below 3 Remove remaining tape liner and press firmly onto glass me...

Страница 7: ...14 16 Edit Time Hours Minutes Seconds 14 16 34 Press OK to accept 01 Feb 18 14 16 Edit Time Hours Minutes Seconds 14 16 34 Press OK to accept 1 Scroll to Edit Time and press the OK button 2 Scroll to...

Страница 8: ...nd select Properties If the File system is identified as FAT32 no formatting is needed Continue to Creating a Wi Fi Configuration File NOTE This view will also show how much space is available on your...

Страница 9: ...file from the computer onto the USB thumb drive Once the file has loaded remove the USB thumb drive from the computer port Connect to Wi Fi Network 1 From the Main Menu select Settings and press the...

Страница 10: ...were last set by the user either during a count or during post count data management 01 Feb 18 14 16 Menu Home Histograms Settings Help 01 Feb 18 14 16 Settings Back Date Time Bluetooth Wifi Gating M...

Страница 11: ...epter 3 0 instrument Do not use sensor if visibly damaged Attach the Sensor 1 When the Scepter 3 0 is powered on it will show Attach the Sensor to begin 2 Holding the sensor with the sensor size numbe...

Страница 12: ...net is recommend use low vacuum 5 Hg Cell Bead and Particle Suspensions Materials Required Cell sample Scepter 3 0 Test Beads Cat No PHCC3BEADS or particle suspensions Note Some particle suspensions...

Страница 13: ...ion and diameter or volume will be displayed on the screen Press the OK button to save up to 999 histograms in the Scepter 3 0 memory and open the Histogram Explorer Press the OK button to return to H...

Страница 14: ...03 Dec 17 Count 0 900 36 24 30 18 12 6 Dia in m 3 342e5 15 685 mL m Conc Dia 01 Feb 18 14 16 Histogram Edit Gates High gate Done 26 575 Low gate 12 585 Count 0 900 36 24 30 18 12 6 Dia in m Switch to...

Страница 15: ...wirelessly Materials Required Wireless Network broadcasting at 2 4 Ghz PC Mac or UNIX Computer connected to the same Wi Fi Network as Scepter 3 0 Scepter 3 0 Instrument Follow Wi Fi Network p 8 direct...

Страница 16: ...MS Press the OK button from any histogram on screen to open the Histogram Explorer menu 01 Feb 18 14 16 Menu Home Histograms Settings Help 01 Feb 18 14 16 Histogram Test 3 03 Dec 17 Count 0 900 36 24...

Страница 17: ...o printer Help Screens Tutorials 1 From the Menu screen scroll to Help press the OK button 2 Scroll to the topic of interest press the OK button 01 Feb 18 14 16 Scepter Software 4 6 01 Feb 18 14 16 Me...

Страница 18: ...ration of cell sample is within recommended operating range See Cell Bead and Particle Suspensions Test sample using Scepter 3 0 Test Beads Cat No PHCC3BEADS to ensure unit operating properly for conc...

Страница 19: ...y needs to be re charged Place Scepter 3 0 on charging stand to recharge the instrument See Battery p 5 Battery is not charging Ensure that the unit is correctly positioned in charging stand An animat...

Страница 20: ...ressing of OK button Fully submerge sensor in sample before pressing OK button Air bubble in sample Mix sample gently to avoid air bubbles Sensor lifted from sample before screen displays Sample Loade...

Страница 21: ...ed on FAT32 formatted USB thumb drive and try again See USB Thumb Drive Formatting p 8 Software update blocked on battery low Insufficient battery to perform update Recharge Scepter 3 0 and try again...

Страница 22: ...ntal Conditions Temperature 15 C to 25 C Maximum Relative Humidity 20 80 non condensing Storage Conditions Temperature 15 C to 30 C Maximum Relative Humidity 20 80 non condensing Scepter 3 0 Sensor Ma...

Страница 23: ...p E GP Sterile Vacuum Filtration System Millipore Express PLUS PES 0 22 500 38 12 SEGPU0538 0 22 500 45 12 SEGPU0545 0 22 1000 38 12 SEGPU1138 0 22 1000 45 12 SEGPU1145 Steritop E GP Sterile Funnel le...

Страница 24: ...Merck KGaA Darmstadt Germany and or its affiliates All Rights Reserved Global Registration Marks of Conformity USA TFB 1003 EUROPE C TAIWAN JAPAN R209 J00212 SOUTH KOREA R C Emd PHCC30000 RUSSIA BELAR...

Отзывы: