
68
Replacing a POD Pak
Basing on
flowrate
One possible reason for a decrease in Milli-Q® Water flowrate is a clogged POD Pak.
The POD Pak should be replaced when it appears to be clogged.
Make sure the POD Pak is not air-locked. Dispense water and open the vent to see if
there is any trapped air. Close the vent after this.
Basing on LCD
message
The POD Pak needs replacement when the following Alert message is displayed.
•
Alert message = REPLACE POD PAK
Placing and
flushing
Follow the instructions delivered with the POD Pak.
Registering
The POD Pak installation has to be registered.
Follow the steps below to register the installation of the POD Pak.
Step
Action
Diagram
1
Start in STANDBY Mode.
2
Select Menu.
Press
.
3
Select Maintenance.
Press
.
Continued on next page
Содержание Millipore
Страница 1: ...1 User Manual Milli Q Reference System...
Страница 2: ......
Страница 6: ......
Страница 37: ...Milli Q Reference System 37 Software Map M a p...
Страница 92: ...Milli Q Reference System FTPF11373 V 2 0 03 2013...