42
Description,
Continued
Set Points
Diagram 1
Diagram 2
Item
Description
Strainer Frequency
Change set points for controlling the
frequency of the message EXAMINE INLET
STRAINER.
Milli-Q® Feed Cond
Change set point controlling the message
MILLI-Q FEED CONDUCTIVITY > SP.
Milli-Q® Inter Res
Change set point controlling the message
MILLI-Q INTER R < SP, PLEASE ORDER Q-
GARD® AND QUANTUM®.
Milli-Q® Product Res
Change set point controlling the message
MILLI-Q RES < SP, REPLACE
Q-GARD® AND QUANTUM®.
Milli-Q® Product TOC
Change set point controlling the message
MILLI-Q TOC > SP.
Millipak®
Change set point controlling the message
REPLACE POD PAK IN XX DAYS (where
1
≤
XX
≤
15).
BioPak®, EDS-Pak®, POD Pak
See above.
Units
Diagram 1
Diagram 2
Item
Description
Milli-Q®
Product
•
Change the displayed units of Milli-Q® Product Water quality.
•
Choices are M
Ω
.cm or
µ
S/cm.
Tank
Volume
•
Change the displayed units of Tank Volume.
•
Choices are % full, Litres or US Gallons.