55
Description,
Continued
Setup
Diagram 1
Diagram 2
Item
Description
Install Date
Change the installation date.
Buzzer
Change the setting for the Buzzer.
MQ Recirc Mode
Change the amount of time that the System
automatically recirculates every hour in READY
Mode.
NOTE
If set to 60 minutes, the daily TOC Curve Check
will not be performed.
POD Flow Stop
Change the amount of time that the POD Unit
dispenses continuously before it automatically
stops.
Temp Comp Mode
Change the Temperature Compensation Mode.
Flow Calibration
Used for performing a flow calibration.
UV 254 nm Activation
Used to activate or deactivate the UV 254 nm
Lamp.
UV 185 nm Activation
Used to activate or deactivate the UV 185 nm
Lamp.
ASM UV Lamp Schedule
•
Used to change the times when the ASM
(Automatic Sanitisation Module) turns on.
•
See the ASM User Manual for more
information.
Network Settings
•
Change Network settings.
•
Contact Millipore SAS for more information.
TOC Curve Check Time
Change the time when a TOC Curve Check is
automatically performed each day.