Registering UV Lamp timers
Introduction
The timer used for each UV Lamp must be reset when the System is installed.
If this is not done, then the message indicating that a Lamp replacement is needed is
shown too early.
The UV Lamp timers need to be reset for:
−
the UV 185 nm Lamp, and
−
the UV 254 nm Lamp.
NOTE:
Before doing this, make sure that the date and time have been checked for accuracy.
Procedure
This procedure shows how to reset the timer used for the UV 185 nm Lamp.
Step
Action
Diagram
1
Place the System in STANDBY
Mode.
2
Select Menu.
Press
.
3
Select Maintenance.
Press
.
4
Select Install UV 185 nm Lamp.
Press
.
Continued on next page