Installation
Version 1.0de - 09/2019
MV-R 100/120
Page 15
1.
Align the supporting surface horizontally.
2.
Place the
MV-R 100/120
on the supporting surface.
$
Make sure the surface is non-slip.
$
Lift and carry the sintering oven only at the unit base.
Electrical connection
Local installation
;
The
MV-R 100/120
requires its own electrical circuit.
;
The building's electrical circuit must have a type-K or type-Z circuit
breaker with a rated current of at least 16 A (other types of circuit
breaker depending on the country of use).
;
An additional residual-current circuit breaker (designed for 30 mA
tripping current) must be installed.
;
To ensure electrically safe operation, the
MV-R 100/120
requires a
protective earth conductor to be connected to the power socket.
;
When selecting the installation location, ensure that the accompanying
mains cable is 2.0 m long. Extending the cable is not permitted. The
supply voltage must be within the rated voltage range of 200-240 V.
Danger
Electricity!
Risk of death from electric shock.
h
Do not touch live cables and components with wet hands.
h
Observe the accident-prevention regulations when working with
electric current.
h
Only connect the unit to a voltage supply that matches the
specifications on the rating plate.