INSTRUCTIONS FOR USE
INSTRUCTIONS FOR USE
|
GB
GB
TECHNICAL DATA
Product no.
Descriptiion
Voltage range
Reader Unit
Power supply
6 V (DC
100-240 V (50–60Hz)
Power consumption
Reader Unit
Power supply
1,4A (DC)
0,25A (50–60Hz)
Operating frequency
133kHz
Pressure measurement range
-66,67 hPa to 3.33 hPa ± 5.32 hPa
Temperature range for pressure measurement
20°C to 45°C
Working distance between antenna and shunt sensor
10 to 30mm
Battery lifetime
min. 5 years
Battery charging cycles
min. 250 charge cycles
Battery self-discharge
Residual charge after 3 months (storage) > 70%
Flammability class of housing
UL 94 HB
Damp protection or impermeability
Reader Unit
Antenna
Power supply
IP44
IP44
IP40
Durability - shock testing
to IEC 60601-1 (3rd edition) 15.3.2
Durability - drop testing
to IEC 60601-1 (3rd edition) 15.3.4.1
Weight:
Reader Unit
Antenna
Power supply
0.600 kg
0.215 kg
0.115 kg
Dimensions (W x H x D)
Reader Unit
Antenna (without cable)
Power supply
144 x 270 x 65mm
100 x 250 x 0.98
55 x 65 x 2.76
Safety rating (Power supply)
II
Conforming to standard
IEC/DIN EN60601-1/ EN 45502-1
EMC
IEC/DIN EN 60601-1-2
AMBIENT CONDITIONS
Operating conditions
Relative humidity
30% to 75%
Ambient
temperature
10°C to 40°C, no con-
densation
Atmospheric pressure
800 hPa to 1.100 hPa
Transport conditions
Relative humidity
15% to 95%
Ambient temperature
0°C to 50°C
Atmospheric pressure
500 hPa to 1.100 hPa
Storage conditions
Relative humidity
15% to 95%
Ambient temperature
10°C to 40°C
Atmospheric pressure
500 hPa to 1 100 hPa
CE-MARKING
Approval to affix the CE marking as defined by
Directive 90/385/EEC for active implantable
medical devices was granted in 2011.
STORAGE AND TRANSPORT
· Store reprocessed products protected from
dust in a dry, dark and cool room which is
exposed to as little bacterial contamination
as possible.
· Ship the product only in its original packa-
ging.
· For storage and transport conditions, see
Chapter Technical Data.
·
Caution: Damage to the product if used
too soon after storage/shipment at tem-
peratures below +10°C.
·
Leave the Reader Unit to acclimatize to
room temperature for about three hours.
DISPOSAL
Adhere to national regulations when
disposing of or recycling the product,
its components and packaging
Products carrying this symbol are
subject to separate collection of electrical and
electronic devices. Within the European Union,
disposal is arranged free of charge by the ma-
nufacturer.
Please contact Christoph Miethke
GmbH & Co. KG if you have any questions con-
cerning the disposal of the product, see Chap-
ter Technical Service.
For recycling of the product please send the
Reader Unit Set for SENSOR RESERVOI
R back
to the manufacturer.