en - Warning and Safety instructions
29
Read the operating instructions carefully before using this holder.
This prevents both personal injury and damage to the rinse fitting.
Keep these operating instructions in a safe place.
Please also read the operating instructions for your washer-
disinfector and pay particular attention to the Warning and Safety
instructions.
The adapter has been approved for the applications stated in these
operating instructions. Applications that go beyond those listed and
use of the adapter with additional nozzles, irrigation sleeves and rinse
fittings from the Miele range of accessories are described in the oper-
ating instructions for the accessories or are to be agreed with Miele.
Conversions, modifications and any other use are not permitted.
New load carriers must be cleaned in the washer-disinfector
without a load prior to first use.
Inspect all mobile units, baskets, modules and inserts daily as de-
scribed in the “Maintenance” section in the Operating instructions for
the washer-disinfector.
For your own safety, wear gloves and, if necessary, a surgical
mask and protective goggles when sorting the instruments and un-
loading the machine. To avoid recontamination, wear clean gloves
when unloading the machine.
To clean and disinfect lumened instruments thoroughly from the in-
side, they must be attached to the relevant holders or nozzles on the
injector manifolds. The connectors must not come loose during a
programme sequence.
At the end of every programme sequence, check that the connectors
have not come loose.
The last rinse, as a minimum, must be carried out with DI water.
Содержание A 865
Страница 2: ...2 de 4 da 15 en 26 es 37 fi 48 fr 59 hr 70 it 81 nl 92 no 103 pt 114 sl 125 cs 136 ...
Страница 146: ......
Страница 147: ......