The conductance measure module has
the same function as the built-in
conductance measure module in the
Aqua Purificator G 7895.
The water conductance is measured
and a quality control of the
demineralized water enabled.
^
The connection of the conductance
measure module is always after the
water demineralization cartridge:
a
water valve
b
non-return valve with pipe venting
c
water demineralized cartridge
d
conductance measure module
f
machine.
^
Install the module at a wall or in a
cabinet.
Do not install the module
on the floor or underneath water
valves.
^
Electrical connection
e
: Plug the
cord into an outlet.
Please also read the separate
Installation Instructions supplied
with this machine.
Conductance measure module