The conductance display shows the
conductance of the demineralized
water, measured in

Z

S/cm (micro

siemens per centimeter) and is a
quality control for the demineralized
water.

If the water demineralization is below
the minimum value, the cartridge needs
to be refilled. A manual is supplied with
the "Original Miele resins replace set
E 316".

Uninstall the cartridge E 318

^

Turn off the water valves.

^

If available, open the water valve for
manual DI water extraction.

^

Open the priming valve the cartridge.

Dry the water exiting from the
cartridge.

.

^

Unscrew the hoses.

^

Turn the cartridge over to let the
residual water run out.

Conductance display

11

Содержание G 7895

Страница 1: ...Operating Instructions Aqua Purificator G 7895 and conductance measuring module C To prevent accidents and machine damage read these instructions before installation or use UV M Nr 06 159 860...

Страница 2: ...tion of the machine 7 Filling Cartridge E 318 with one time resins E 315 8 Connection of E 318 cartridge 9 Connection 10 Conductance display 11 Conductance measure module 12 Electrical connection 13 A...

Страница 3: ...a Machine lid b Door handle c Conductivity display meter d Power cord e Entrance hose blue connects to tap water f Hose H2O pur green connects to labwasher Guide to the machine 3 1 2 3 4 5 6...

Страница 4: ...g Water demineralization cartridge h Connection Exit i Connection Entrance Guide to the machine 4 EINGANG ENTREE ENTRANCE 7 8 9...

Страница 5: ...hen maintained only by a Miele Authorized Service Technician Do not install the machine in an area where a danger of explosion or freezing may be present Be certain this appliance is properly installe...

Страница 6: ...hydrochloric acid do not come into contact with the steel outer casing of the machine Using accessories Only specific equipment made by Miele should be connected to this machine Consult Miele on the...

Страница 7: ...ns The resins are made for one time use only Dispose them accordingly To ease the emptying and refilling of the water demineralization cartridge E 318 we recommend as additional equipment to the cartr...

Страница 8: ...t Open the lid Filling the resins into the cartridge Insert the funnel into the cartridge opening Empty both bags of resins into the cartridge Clean the cartridge opening of any resins Insert the lid...

Страница 9: ...hine front Unscrew the caps from the cartridge connection Screw the hoses onto Entrance blue to Entrance H2O pur green to Exit See graphic for hose location Connection of E 318 cartridge 9 EINGANG ENT...

Страница 10: ...ceed 68 F 20 C Plug the machine into the outlet to operate the conductance display Priming the cartridge Open the venting valve arrow by turning it counterclockwise Open the cold water valve until wat...

Страница 11: ...minimum value the cartridge needs to be refilled A manual is supplied with the Original Miele resins replace set E 316 Uninstall the cartridge E 318 Turn off the water valves If available open the wat...

Страница 12: ...module is always after the water demineralization cartridge a water valve b non return valve with pipe venting c water demineralized cartridge d conductance measure module f machine Install the modul...

Страница 13: ...he meter Pease contact Miele Technical Service at the phone numbers located on the back cover of this manual is there are any questions To increase security it is recommended to install a protective s...

Страница 14: ...our authorized Service Dealer or the Miele Service Department See back cover for addresses Please quote the model and serial number of your machine This information can be found on the data plate see...

Страница 15: ...of safely and kept out of the reach of children Danger of suffocation Disposal of the appliance Appliances contain materials which can be recycled Please contact your local authorities about recyclin...

Страница 16: ...M Nr 06 159 860 00 All rights reserved 1104 This paper has been bleached without the use of chlorine...

Отзывы: