Fit the bellows onto the E 431
injector jet.

A special combination of E 464 and
E 454 jets (optional extra) is required
for processing laryngeal masks.

^

Remove the height adjustment spring
from injector jet E 454.

^

Screw injector jet E 454 onto the
E 464 holder.

^

Now screw the jet combination into
the position released by one of the
E 466 injector jets and tighten with a
7 mm spanner.

^

Expel as much air as possible from
the laryngeal mask and the balloon
lead.

^

Attach the laryngeal mask to the jet
combination.

^

Attach the balloon lead with the valve
to the spring clip.

^

Arrange small items, such as
Y-pieces, couplings or connectors, in
insert E 507.

^

Lightweight items should be secured
to prevent them blocking the spray
arm or being caught up with the
magnetic strip on the automatic
mobile unit recognition system.

If bellows are to be processed, one of
the E 466 injector jets must be replaced
with a special E 431 injector jet for
bellows which has a larger support
ring.

To do this:

^

Use a 7 mm spanner to unscrew one
of the E 466 injector jets.

^

Then screw the E 431 injector jet into
the vacated position and tighten with
the spanner.

To process laryngeal masks with two
tubes, exchange one of the injector jets
with a spring clip for a short injector jet
(supplied separately). To do this:

^

Using a 9 mm screwdriver, unscrew
the injector jet with the spring clip.

^

Screw the short injector jet into the
holder, and tighten with the
screwdriver.

^

Push the cone and spring clip up the
long injector jet so that the longer
tube on the laryngeal mask on the
short injector jets is not kinked.

^

Expel as much air as possible from
the laryngeal mask and the balloon
lead.

^

Attach the laryngeal mask to the two
injector jets.

^

Attach the balloon lead with the valve
to the spring clip.

,

After processing, check that

items are airtight and function
correctly.