2) Extending the input leads may introduce more noise during measurement.

If extra extension lead is necessary, please ensure the extension leads

are of equal length and the input signal frequency is under 10MHz.

Otherwise, measurement errors may occur.

Warranty

6

Micsig warrants that this probe will be free from defects in materials and 

workmanship for a period of one (1) year and accessories for a period of 

six(6) months from the date of shipment. If any such product proves 

defective during this warranty period, Micsig, at its option, either will repair

the defective product without charge for parts and labor, or will provide a 

replacement  in  exchange  for  the  defective  product.  Parts,  modules 

and replacement products used by Micsig for warranty work may be new 

or reconditioned to like new performance. All replaced parts, modules and 

products become the property of Micsig.

Shenzhen Micsig Instruments Co., Ltd.

Add: 1F, Bldg A, Huafeng International Robot Industrial Park, 

Hangcheng Rd, Bao'an District, Shenzhen, Guangdong, China, 518126

WEB: www.micsig.com

TEL: +86-755-88600880

E-mail: [email protected]

3) While measuring a high frequency signal, don't touch the end of the

input lead with your hand or other objects. Otherwise, it may affect

the accuracy of the measurement.

4) Ensure that you use an oscilloscope with an input impedance of at

least 1M

Ω a

nd bandwidth of at least 100MHz.

5) Turn on the oscilloscope or externally powered instrument and let

the probe and equipment warm up for 20 minutes.

Best Practices

5

1) Twisting the input leads together can help reduce noise and improve

the probe's high frequency response when measuring signals. Please

view the diagram below for an example:

6) When a substantial change in temperature or other circumstances affect

the accuracy of the probe's zero point, a calibration is needed: short the

input terminals of the probe, then power the probe while simultaneously

pressing the Attenuation Range Buttons/Indicators for three seconds.

Electronics Source 256 5th - 6th Fl.,Phahonyothin Rd.,Samsennai, Phayathai,Bangkok 10400

Содержание DP10007

Страница 1: ...To avoid an electric shock all devices that require grounding must be connected to earth ground Before making connections to the input or output terminals of the probe ensure that the test instrument is properly grounded if necessary Measurement safety Always be aware of the voltage rating of the probe and the measurement accessories you are using and of the maximum amplitude of the signal you int...

Страница 2: ...asurement voltage can reach up to 700V 1300V 5600V 1 2 Specification Operating Humidity Operating Temperature 15mVrms 10X 8MΩ 1 25pF differential 4MΩ 2 5pF single ended DP10007 100MHz 3 5ns 10X 100X 1 70V 10X 700V 100X 80dB 50Hz 50dB 1MHz 7V Button light flashes DC 5V USBSupply 1 25W 14 5cm 6cm 2 7cm Approx45cm Approx90cm 0 40 10 85 CAT II 600V CAT I 1000V 60dB 20KHz 40dB 10MHz 60mVrms 100X to gro...

Страница 3: ...illoscope Connect the probe output BNC to the oscilloscope channel input Note make sure the oscilloscope is properly grounded if necessary 3 Set the appropriate attenuation range according to the measured voltage 4 Connecting the input to the device under test Using the appropriate input accessory connect the probe to the device under test to start the measurement If the attenuation range indicato...

Страница 4: ...uafeng International Robot Industrial Park Hangcheng Rd Bao an District Shenzhen Guangdong China 518126 WEB www micsig com TEL 86 755 88600880 E mail sales micsig com 3 While measuring a high frequency signal don t touch the end of the input lead with your hand or other objects Otherwise it may affect the accuracy of the measurement 4 Ensure that you use an oscilloscope with an input impedance of ...

Отзывы: