1. Connect the probe BNC cable to the channel interface of the oscilloscope;
2. Power the probe control module with the standard USB power adapter provided
with the probe.
Caution: Must use the adapter provided with the probe for power supply, DO NOT use
USB port on the oscilloscope.
Installation
Caution: The probe head is a precision part. Do not drop the probe or physically shock
it, twist it, or drastically alter its surroundings. Do not insert wires larger than 5.0 mm in
diameter into the probe grip, as this may damage the probe.
Connecting to an oscilloscope
Degaussing and Zero
Note:
Before degaussing and zero the probe, be sure to verify that the probe head is
locked.
When the oscilloscope recognizes the probe for the first time, the probe will
automatically degauss and zero.
To ensure accurate measurements, degauss the probe under following conditions:
* Turn on the measuring system and allow a 20-minute warm-up time
* Before connecting the probe to the wire
* Whenever there is a current or thermal overload
* Whenever the probe is placed in a strong external magnetic field
Press the button "Zero
□
", the button lights up, probe will conduct degaussing and zero
calibration. If the button light flashes 5 times, it means zero calibration failed.
Installation