SimpliFiber
ENGLISH - 24
press
.
4. Power on
SimpliFiber Meter
. The unit will auto-
sense the appropriate wavelength.
Note: When using a light source other than
SimpliFiber
Source, press the
to select the
appropriate wavelength. Both units must be set
to the same wavelength.
5. Press, hold and release the
button to set
and save the new
reference value.
The reference
value will be
displayed.
6. Disconnect the Launch Cable from the
SimpliFiber
Meter.
Measuring Loss using Method B
When Method B was used to set the reference,
proceed as follows:
1. Do not disconnect the launch cable from the
SimpliFiber
Source unit.
2. Connect the fiber to be measured between the
SimpliFiber Meter
output and the launch cable that
is attached to
SimpliFiber
Source. An extra