61
(1)
V ramp Time
Sets the voltage increase time for the DC insulation and AC tolerance test. The increase time for the
voltage of the DC insulation test can be set between 0~64 phases, the smaller the setting value the faster it
increases.
(2)
Arc Detection
Sets whether to enable arc detection.
(3)
Single Node
Sets whether to test for empty pins; it is mainly used to test tests with no circuits.
(4)
DC arc Sensitivity
Sets the DC arc detection sensitivity 1-8; the lower the value the more sensitive it is, the higher the value
the less sensitive it is.
(5)
AC arc Sensitivity
Sets the AC arc detection sensitivity 1-8; the lower the value the more sensitive it is, the higher the value
the less sensitive it is.
(6)
A (B, C, D, E, F, G, H) PORT Final Pin
Sets the final pin of A (B, C, D, E, F, G, H)
PORT.
Binary Range Number
1-14, used with the DUT quantity option in test environment setting.
Binary Range
Sets the range of the DUT.