41
(2)
Maximum allowable capacity (MAX. Capacitance):
Can be set between 0.05u~1u. This value
should be set to a great number when the cables are longer in order to have stable tests.
(3)
Short circuit pin determination (IDENTIFY O/S ENDS)
: Determine which pins are open circuit
or short circuit.
(4)
Unilateral test sensitivity (SINGLE SIDE SENDS.)
: The lower the value the more sensitive it is.
0~999.
(5)
Instant short circuit test time (INTER. O/S Time):
0~60s.
(6)
Instant short circuit test maximum pin (INTER. O/S Scan)
: Determined according to model.
(7)
TYPE C voltage (TYPE C Volt.):
TYPE C test voltage 0.02v~4v.
(8)
TYPE C fixture (TYPE C Fix.):
Uses standard fixture test.
2.
Cable type: {Normal}
Press [Learn], the LCD screen will display as follows: