7.9 Exchanging the Specimen Clamping Systems
Exchanging the specimen clamping system
Release (press downwards) the eccentric cam lever.
Withdraw from the left out of the specimen orientation device.
Screw out the orientation screws and spring loaded pin little way.
Withdraw the specimen clamping fixture from the front from the specimen orientation
device.
Important!
After exchanging the clamping system, make sure that the spring
loaded pin is screwed in again, otherwise there will be insufficient
spring pressure imparted.
Note
In order to increase leverage at the handle of the eccentric cam
lever, the handle sleeve can be pulled out. After releasing, the handle
sleeve springs back into its
nominal position.