18
Host sends: 01 03 00 03 00 02 34 0B
Slave response: 01 03 04
D0 D1 D2 D3
CRCL CRCH
d) Positive total decimal place
message
definition
(
FLOAT
)
Host sends: 01 03 00 05 00 02 D4 0A
Slave response: 01 03 04
D0 D1 D2 D3
CRCL CRCH
e) Positive total unit message definition
(
SHORT/ List type
,
same as reverse unit
message
)
Host sends: 01 03 00 07 00 01 35 CB
Slave response: 01 03 04
D0
CRCL CRCH
D0 Definition
:
D0
0
1
2
3
4
5
unit
L
m
3
gal
igal
kg
t
f)
Reverse total integer low 4 bit
message
definition
(
LONG
)
Host sends: 01 03 00 08 00 02 45 C9
Slave response: 01 03 04
D0 D1 D2 D3
CRCL CRCH
g) Reverse total decimal places
message
definition
(
FLOAT
)
Host sends: 01 03 00 0A 00 02 E4 09
Slave response: 01 03 04
D0 D1 D2 D3
CRCL CRCH
h) Reverse total unit
message
definition
(
SHORT/
List type
)
Host sends: 01 03 00 0C 00 01 44 09
Slave response: 01 03 04
D0
CRCL CRCH
D0 Definition
:
D0
0
1
2
3
4
5
unit
m
3
L
t
kg
gal
igal
i)
Excitation alarm
(
SHORT
)
Host sends: 01 03 00 0D 00 01 15 C9
Slave response: 01 03 02
00 T/F
CRCL CRCH
j)
Electrode alarm
(
SHORT
)
Host sends: 01 03 00 0E 00 01 E5 C9
Slave response: 01 03 02
00 T/F
CRCL CRCH
k) Empty pipe alarm
(
SHORT
)
Host sends: 01 03 00 0F 00 01 B4 09
Slave response: 01 03 02
00 T/F
CRCL CRCH
l)
Upper limit alarm
(
SHORT
)
Host sends: 01 03 00 10 00 01 85 CF
Slave response: 01 03 02
00 T/F
CRCL CRCH
m) Lower limit alarm
(
SHORT
)