Gemini VII
Appendix C
239-42828-01 - Dec 2012
C-1
C. CALCULATIONS
This appendix contains the calculations used in the analysis program.
Saturation Pressure
Saturation pressure (Po) is selected on the Po and Temperature Options dialog. It may be entered or
measured in the Po tube or on port 3.
• If entered, Po = user-entered value
• If measured, Po = equilibrated pressure reading after saturating Po tube or the sample tube
Relative Pressure Calculations
Po
I
Po
1
Po
2
Po
1
–
t
I
to
1
–
to
2
to
1
–
----------------------------------------------------------
+
=
If Po was measured at intervals in the Po tube and several data points were collected in the interval:
• For points taken before the previous Po measurement,
• For points taken after the previous Po, use the previous Po:
Po
I
= Po
1
If Po was entered, use the entered Po:
Po
I
= Po
E
P
rel
1
P
1
Po
I
--------
=
Calculate relative pressure for the I
th
data point:
where
Po
I
=
saturation pressure for the I
th
data point (mmHg)
Po
1
=
most previous measured saturation pressure before I
th
data point (mmHg)
Po
2
=
first measured saturation pressure after the I
th
data point (mmHg)
Po
C
=
calculated Po
Po
E
=
entered saturation pressure (mmHg)
t
I
=
time when the I
th
data point was taken (minutes)
to
1
=
time when Po
1
was measured (minutes)
to
2
=
time when Po
2
was measured (minutes)
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...