ASAP 2420 1-mmHg Transducer Option
Installing the 1-mmHg Transducer and Micropore Options
Feb 2010
15
16. Install the transducer on top of the transducer block; tighten securely with a wrench.
17. Connect the ribbon cable to the 1-mmHg transducer and to
J7
on the transducer board
(202-17706-001).
18. Install the 10-mmHg transducer, with new front and rear ferrules, in the front of the
transducer block. Use the connector nut you removed earlier; tighten the nut securely with
a wrench.
Transducer block
Connector nut (between
transducer and block)
J7
1-mmHg transducer
Transducer board
10-mmHg transducer Connector nut
Use new ferrules from the upgrade kit. Do not replace the ones you removed, they may
be damaged.