BP W3 Comfort

EN

This device may only be used for the purposes described in these 

instructions. The manufacturer cannot be held liable for damage 

caused by incorrect application.

This device comprises sensitive components and must be treated 

with caution. Observe the storage and operating conditions 

described in the «Technical Specifications» section.

Protect it from:

- water and moisture

- extreme temperatures

- impact and dropping

- contamination and dust

- direct sunlight

- heat and cold

The cuffs are sensitive and must be handled with care.

Only pump up the cuff once fitted.

Do not use this device close to strong electromagnetic fields such

as mobile telephones or radio installations. Keep a minimum 

distance of 3.3 m from such devices when using this device.

Do not use this device if you think it is damaged or notice anything 

unusual.

Never open this device.

If the device is not going to be used for a prolonged period the 

batteries should be removed.

Read the additional safety information provided within the indi-

vidual sections of this instruction manual.

The measurement results given by this device is not a diagnosis. 

It is not replacing the need for the consultation of a physician, 

especially if not matching the patient’s symptoms. Do not rely on 

the measurement result only, always consider other potentially 

occurring symptoms and the patient’s feedback. Calling a doctor 

or an ambulance is advised if needed.

Ensure that children do not use this device unsupervised; 

some parts are small enough to be swallowed. Be aware of 

the risk of strangulation in case this device is supplied with 

cables or tubes.

Device care

Clean the device only with a soft, dry cloth.

Cleaning the cuff

Carefully remove any marks on the cuff with a damp cloth and mild 

detergent.

Accuracy test

We recommend this device is tested for accuracy every 2 years or 

after mechanical impact (e.g. being dropped). Please contact your 

local Microlife-Service to arrange the test (see foreword).

Disposal

Batteries and electronic devices must be disposed of in accor-

dance with the locally applicable regulations, not with 

domestic waste. 

Содержание W3 Comfort

Страница 1: ...olife com Asia Microlife Corporation 9F 431 RuiGuang Road NeiHu Taipei 11492 Taiwan R O C Tel 886 2 8797 1288 Fax 886 2 8797 1283 Email service microlife com tw www microlife com North Central South America Microlife USA Inc 1617 Gulf to Bay Blvd 2nd Floor SuiteA Clearwater FL33755 USA Tel 1 727 442 5353 Fax 1 727 442 5377 Email msa microlifeusa com www microlife com C C C ...

Страница 2: ...arihi Fecha de compra Data da compra Datum van aankoop Ημερομηνία αγοράς Specialist Dealer Revendeur Categoria riven ditore Fachhändler Uzman satıcı Distribuidor especializado Revendedor autorizado Specialist Dealer Εξειδικευμένος αντιπρόσωποςa اﻟﻜﻔﺎﻟﺔ ﺑﻄﺎﻗﺔ اﻟﺨﻠﻔﻲ اﻟﻐﻄﺎء اﻧﻈﺮ اﻟﻜﻔﺎﻟﺔ ﺑﻄﺎﻗﺔ اﻟﻤﺸﺘﺮي اﺳﻢ اﻟﻤﻮدﻳﻞ اﻟﺘﺴﻠﺴﻞ رﻗﻢ اﻟﺸﺮاء ﺗﺎرﻳﺦ اﻟﻤﺨﺘﺺ اﻟﺘﺎﺟﺮ اﻟﻜﻔﺎﻟﺔ ﺑﻄﺎﻗﺔ اﻟﺨﻠﻔﻲ اﻟﻐﻄﺎء اﻧﻈﺮ اﻟﻜﻔﺎﻟﺔ ﺑﻄﺎﻗﺔ اﻟ...

Страница 3: ...rs or older Dear Customer Thisdevicewasdevelopedincollaborationwithphysiciansandclinicaltests carried out prove its measurement accuracy to be of a very high standard If you have any questions problems or want to order spare parts please contact your local Microlife Customer Service Your dealer or pharmacy will be able to give you the address of the Microlife dealer in your country Alternatively v...

Страница 4: ...replacement Which batteries and which procedure Using rechargeable batteries 6 Error Messages 7 Safety Care Accuracy Test and Disposal Safety and protection Device care Cleaning the cuff Accuracy test Disposal 8 9 Guarantee Technical Specifications 1 Using the Device for the First Time Inserting the batteries After you have unpacked your device first insert the batteries The battery compartment 5 ...

Страница 5: ... height as your heart 3 Taking a Blood Pressure Measurement 1 Press the ON OFF button 1 to start the measurement 2 The cuff fit check AO on the display indicates that the cuff is perfectly placed If the icon AO A appears the cuff is fitted subop timally but it is still ok to measure 3 The cuff will now pump up automatically Relax do not move and do not tense your arm muscles until the measurement ...

Страница 6: ...tically stores the last 30 measurement values Viewing the stored values Press the M button 3 briefly when the device is switched off The display first shows M AK and A which stands for the average of all stored values Pressing the M button again displays the previous value Pressing the M button repeatedly enables you to move from one stored value to another Blood pressure readings with suboptimal ...

Страница 7: ...replacement batteries Flat battery replacement When the batteries are flat the battery symbol AQ will flash as soon as the device is switched on flat battery displayed You cannot take any further measurements and must replace the batteries 1 Open the battery compartment 5 by pulling off the cap 2 Replace the batteries ensure correct polarity as shown by the symbols in the compartment 3 To set date...

Страница 8: ...scription Potential cause and remedy ERR 1 AN Signal too weak The pulse signals on the cuff are too weak Re position the cuff and repeat the measurement ERR 2 AO Error signal During the measurement error signals were detected by the cuff caused for instance by movement or muscle tension Repeat the measurement keeping your arm still ERR 3 AO No pressure in the cuff An adequate pressure cannot be ge...

Страница 9: ...d Read the additional safety information provided within the indi vidual sections of this instruction manual The measurement results given by this device is not a diagnosis It is not replacing the need for the consultation of a physician especially if not matching the patient s symptoms Do not rely on the measurement result only always consider other potentially occurring symptoms and the patient ...

Страница 10: ...12 g including batteries Dimensions 81 x 65 x 21 mm Cuff size 13 5 23 cm 5 25 9 05 inches Measuring procedure oscillometric corresponding to Korotkoff method Phase I systolic Phase V diastolic Measurement range 20 280 mmHg blood pressure 40 199 beats per minute pulse Cuff pressure display range Resolution Static accuracy Pulse accuracy Voltage source Battery lifetime IP Class Reference to standard...

Отзывы: