background image

12

FR1MF1

EN

Memory:

Backlight:

Ambient
temperature:
Storage
temperature:
Automatic
Switch-off:

Battery
Dimensions
Weight:
Reference to
standards:

The stipulations of EU-Directive 93/42/EEC for Medical Devices 

Class  IIa have been fulfilled. 

Technical alterations reserved.
According to the Medical Product User Act a biennial technical 
inspection is recommended for professional users.
Please observe the applicable disposal regulations.

15.www.microlife.com

Detailed user information about our thermometers and blood pressure 
monitors as well as services can be found at www.microlife.com.

    3 short beeps

• 30 readings recall in the Memory Mode with

    time and date record

• The display light will be GREEN for 4 second

   

   when the unit is turned ON.

•  The display light will be GREEN for 5 seconds,    

  when a measurement is completed with a 
  reading less than 37.5 °C (99.5 °F).

•  The display light will be RED for 5 second, 

when a measurement is completed with a 
reading equal to or higher than 37.5 °C(99.5 
°F). 

Body Mode: 16-40.0 °C (60.8-104.0 °F)
Object Mode: 5-40.0 °C (41-104.0 °F)                 
-20 °C to 50 °C (-4 °F to 122 °F) 
15-95 % relative maximum humidity
Approx. 1 minute after last measurement

has been taken.
2 x 1.5 V Batteries; size AAA
141.1 × 43.3 ×36.9 (mm )
90 g (with battery), 67 g (w/o battery)

ASTM E1965;
IEC 60601-1; IEC 60601-1-2 (EMC)

Содержание FR1MF1

Страница 1: ...Microlife AG Espenstrasse 139 9443 Widnau Switzerland www microlife com IB FR1MF1 EN 2113...

Страница 2: ...Microlife NON CONTACT FR1MF1 EN FR1MF1 3 sec 3 sec...

Страница 3: ...Read the instructions carefully before using this device Microlife FR1MF1...

Страница 4: ...Microlife FR1MF1...

Страница 5: ...rement complete AM Body Mode AN Object Mode AO Low Battery Indicator AP Changing between Celsius and Fahrenheit AQ Recall Mode AR Recall the last 30 Readings AS Measured Temperature too high BT Measur...

Страница 6: ...nt The instrument performs a self test every time it is switched on to always guarantee the specified accuracy of any measurements This Microlife Thermometer is intended for the periodic measurement a...

Страница 7: ...es of this Thermometer Measurements in a matter of seconds The innovative infrared technology allows measurement without even touching the object This guarantees safe and sanitary measurements within...

Страница 8: ...Alarm 10 short beeps and a red LCD backlight alert the patient that he she may have a temperature equal to or higher than 37 5 C 2 Important Safety Instructions This instrument may only be used for t...

Страница 9: ...a prolonged period the batteries should be removed 3 How this Thermometer measures Temperature This thermometer measures infrared energy radiated from the forehead as well as objects This energy is c...

Страница 10: ...en set the day Follow the instructions above to set the day 12 or 24 hours System mode hours and minutes Once you have set the minutes and pressed the M button 8 the date and time are set and the time...

Страница 11: ...enter of the forehead with a distance of no more than 5 cm Please remove any hair sweat or dirt from your forehead before measuring to improve the accuracy of the readings 4 Press the START button4and...

Страница 12: ...mperature readings may vary according to location Doctors recommend rectal measurement for newborn infants within the first 6 months as all other measuring methods might lead to ambiguous results If u...

Страница 13: ...n C and F by pressing the START button 4 When the measurement scale has been chosen wait for 5 seconds and the unit will automatically enter the ready for measuring mode 9 How to recall 30 readings in...

Страница 14: ...orrectly Also check polarity and of the batteries Flat battery indication BO If this icon is the only symbol shown on the display the batteries should be replaced immediately 11 Cleaning and Disinfect...

Страница 15: ...d batteries accidents or non compliance with the operating instructions Please contact Microlife service 14 Technical Specifications Type Measurement range Resolution Measurement accuracy Display Acou...

Страница 16: ...ervices can be found at www microlife com 3 short beeps 30 readings recall in the Memory Mode with time and date record The display light will be GREEN for 4 second when the unit is turned ON The disp...

Страница 17: ...13 FR1MF1 Guarantee Card Name of Purchaser Serial Number Date of Purchase Specialist Dealer...

Страница 18: ...14 FR1MF1 EN...