background image

7

will indicate whether the beeper is activated by displaying a 
beeper icon on the screen.

If no button is pressed for 5 seconds, the device automatically 
switches into the ready for measurement mode.

6. Changing between Body and Object Mode

For changing from body to object mode, slide the mode Switch 

6

  at the side of the thermometer upwards.For switching back to 

body mode, slide the switch down again.

7. Directions for Use

Measuring in Body Mode
1. Press the ON/OFF button 

5

. The display 

3

 is activated to 

    show all segments for 1 second.
2. When the «°

C

» or «°

F

» icon is flashing, a beep sound is heard 

and the thermometer is ready for measuring 

AK

.

3. Aim the thermometer at the center of the forehead with a 

   distance of no more than 5 cm. 

Please remove any hair, 

sweat or dirt from your forehead before measuring to improve 
the accuracy of the readings.

4.

 Press the START button

4

and steadily move the ther-

mometer

 

from the middle of the forehead to the temple area 

(about 1 cm above the eyebrow). The activated blue tracking 
light will indicate the area being measured. After 3 seconds 
a long beep will verify the completion of measurement. If the 
temple 

    area hasn’t been reached before the long beep is heard, 

   repeat he measurement as described above, but move the 

thermometer a little faster.

5. Read the recorded temperature from the LCD display.
Measuring in Object Mode
1. Follow steps 1-2 above, then aim the thermometer at the center 

of the object you want to measure with a distance of no more 
than 5 cm. Press the START button 

4

 After 3 seconds a long 

Содержание FR1MF1

Страница 1: ...Microlife AG Espenstrasse 139 9443 Widnau Switzerland www microlife com IB FR1MF1 EN 2113...

Страница 2: ...Microlife NON CONTACT FR1MF1 EN FR1MF1 3 sec 3 sec...

Страница 3: ...Read the instructions carefully before using this device Microlife FR1MF1...

Страница 4: ...Microlife FR1MF1...

Страница 5: ...rement complete AM Body Mode AN Object Mode AO Low Battery Indicator AP Changing between Celsius and Fahrenheit AQ Recall Mode AR Recall the last 30 Readings AS Measured Temperature too high BT Measur...

Страница 6: ...nt The instrument performs a self test every time it is switched on to always guarantee the specified accuracy of any measurements This Microlife Thermometer is intended for the periodic measurement a...

Страница 7: ...es of this Thermometer Measurements in a matter of seconds The innovative infrared technology allows measurement without even touching the object This guarantees safe and sanitary measurements within...

Страница 8: ...Alarm 10 short beeps and a red LCD backlight alert the patient that he she may have a temperature equal to or higher than 37 5 C 2 Important Safety Instructions This instrument may only be used for t...

Страница 9: ...a prolonged period the batteries should be removed 3 How this Thermometer measures Temperature This thermometer measures infrared energy radiated from the forehead as well as objects This energy is c...

Страница 10: ...en set the day Follow the instructions above to set the day 12 or 24 hours System mode hours and minutes Once you have set the minutes and pressed the M button 8 the date and time are set and the time...

Страница 11: ...enter of the forehead with a distance of no more than 5 cm Please remove any hair sweat or dirt from your forehead before measuring to improve the accuracy of the readings 4 Press the START button4and...

Страница 12: ...mperature readings may vary according to location Doctors recommend rectal measurement for newborn infants within the first 6 months as all other measuring methods might lead to ambiguous results If u...

Страница 13: ...n C and F by pressing the START button 4 When the measurement scale has been chosen wait for 5 seconds and the unit will automatically enter the ready for measuring mode 9 How to recall 30 readings in...

Страница 14: ...orrectly Also check polarity and of the batteries Flat battery indication BO If this icon is the only symbol shown on the display the batteries should be replaced immediately 11 Cleaning and Disinfect...

Страница 15: ...d batteries accidents or non compliance with the operating instructions Please contact Microlife service 14 Technical Specifications Type Measurement range Resolution Measurement accuracy Display Acou...

Страница 16: ...ervices can be found at www microlife com 3 short beeps 30 readings recall in the Memory Mode with time and date record The display light will be GREEN for 4 second when the unit is turned ON The disp...

Страница 17: ...13 FR1MF1 Guarantee Card Name of Purchaser Serial Number Date of Purchase Specialist Dealer...

Страница 18: ...14 FR1MF1 EN...

Отзывы: