19
Err 3
If the cuff takes too long to inflate, it is possible that the cuff is incorrectly positioned
or that the connection with the hose is not airtight.
Check the connections and repeat the measurement.
Err 5
An average value could not be calculated due to unstable conditions while the
readings were being taken.
Repeat the measuring process, keeping your arm still and not speaking.
Err 6
There were too many errors during the measurement in MAM mode, making it impossi-
ble to obtain a final result.
Read through the checklist for performing reliable measurements and then repeat the
measurement.
HI
The pressure in the cuff is too high (over 300 mmHg) or the pulse is too high (over 200
beats per minute).
Relax for 5 minutes and repeat the measurement.
LO
The pulse is too low (less than 40 beats per minute).
Repeat the measurement.
4.5. Switching off the unit
If no buttons are pressed for approx. 1 minute, the device automatically switches to
standby mode. You can also switch to standby mode by pressing the
START
button
w
.
The blood pressure meter functions are now switched off. The time and the selected
user memory are continuously shown on the display.
The alarm function remains active. If you do not want to use the time display and the
alarm function, you can also completely switch off the device by sliding the
ON/OFF
switch
V
on the back of the device into the “
OFF
” position.
4.6. IOS & Android BT APPs instruction
Compatibility:
- Paring your device running iOS 6.0 or better:
iPhone, iPad, and iPod touch.
- Paring your device running Android 2.3.3 and Later:
Android Phone, Android Tablets
2G-IB-edit for NF1-2B_rev_2.3:MLU3AC1-JAN05-2 14-11-20 10:25 Page 19