7. Error messages/troubleshooting
If an error occurs during a measurement, the measurement is
discontinued and a corresponding error code is displayed
(Example: Error No. 2).
Error No.
Possible cause(s)
ERR 1
No pulse was detected.* Ensure the cuff is properly placed
on the inside of your wrist and repeat a measurement.
ERR 2
Unnatural pressure impulses influenced the measurement
result. Reason: The arm was moved during the measurement
(artifact). Repeat, keeping still and quiet.
ERR 3
Inflation of the cuff takes too long. The cuff is not correctly
seated. Re-position cuff and repeat the measurement.
ERR 5
The measured readings indicated an unacceptable difference
between systolic and diastolic pressures. Take another read-
ing following directions carefully. Contact your doctor if you
continue to get unusual readings.
HI
The cuff pressure is too high. Relax for 5 minutes and repeat
the measurement.*
LO
The pulse is too low (less than 40). Repeat the measurement.*
* If this or any other problem occurs repeatedly, please consult
your doctor.