5.5. Inflating further
If the cuff is not inflated sufficiently, the
measurement is stopped after a few seconds and a
flashing arrow pointing upwards appears in the
display.
It is then necessary to inflate at least 20 mmHg
higher than the previous value.
The instruction to inflate further can appear several
times if the pressure is still not sufficient.
Attention:
If the cuff is inflated too far (over 300 mmHg), «HI»
appears in the display and a warning signal is
sounded. The cuff pressure must then be
immediately released using the quick-release valve
at the pump ball!
As soon as the cuff is inflated sufficiently and the
measurement is proceeding, continue as described
under points c + d.
5.6. Discontinuing a measurement
If it is necessary to interrupt a blood pressure
measurement for any reason, press the quick
release valve at the pump ball. The device
immediately lowers the cuff-pressure automatically.
5.7. Memory – displaying the last
measurement
This blood pressure monitor automatically stores the
last measurement value. By pressing the MEMORY
button for at least 3 seconds, the last measurement
will be displayed.
☞
Further information
Measurements should not occur soon after each other, to ensure accurate readings. Wait at
least
1 minute
in a relaxed position, sitting or lying, before you repeat a measurement.