a result of low use of the instrument, even when switched off).
Always remove the rechargeable batteries if you do not intend
to use the instrument for a week or more.
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.
11. Using a Mains Adapter
You can operate this instrument using the Microlife mains adapter
(DC 6V, 600mA).
Only use the Microlife mains adapter available as an original
accessory appropriate for your supply voltage.
Ensure that neither the mains adapter nor the cable are
damaged.
1. Plug the adapter cable into the mains adapter socket
4
in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is
consumed.
12. Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
* Please consult your doctor, if this or any other problem occurs
Error
Description Potential cause and remedy
«ERR 1» Signal too
weak
The pulse signals on the cuff are too
weak. Re-position the cuff and repeat
the measurement.*
«ERR 2»
BS
Error signal During the measurement, error signals
were detected by the cuff, caused
for instance by movement or muscle
tension.
Repeat the measurement, keeping your
arm still.
«ERR 3»
CT
No pressure
in the cuff
An adequate pressure cannot be
generated in the cuff. A leak may have
occurred. Check that the cuff is correctly
connected and is not too loose. Replace
the batteries if necessary. Repeat the
measurement.
«ERR 5» Abnormal
result
The measuring signals are inaccurate
and no result can therefore be displayed.
Read through the checklist «Section 4.»for
performing reliable measurements and
then repeat the measurement.*
«ERR 6» AFIBsens
TM
/ MAM Mode
There were too many errors during the
measurement in AFIBsens
TM
/ MAM
mode, making it impossible to obtain a
final result. Read through the checklist
«Section 4.»
for performing reliable
measurements and then repeat the
measurement.*
«HI»
Pulse or cuff
pressure too
high
The pressure in the cuff is too high (over
300 mmHg) OR the pulse is too high
(over 200 beats per minute). Relax for 5
minutes and repeat the measurement.*
«LO»
Pulse too
low
The pulse is too low (less than 40 beats
per minute). Repeat the measurement.*