EN

A7 Touch BT

measure in the future. The arm with the higher blood pressure 

should be measured.

4.  Remove close-fitting garments from the upper arm. To avoid 

constriction, shirt sleeves should not be rolled up - they do not 

interfere with the cuff if they are laid flat.

5. Always ensure that the correct cuff size is used (marking on the 

cuff).

  • Fit the cuff closely, but not too tight.

  • Make sure that the cuff is 2 cm above your elbow.

  • The artery mark located on the cuff ( 3 cm long bar) must lie 

over the artery which runs down the inner side of the arm.

  • Support your arm so it is relaxed.

  • Ensure that the cuff is at the same height as your heart.

6.  Slide the lock switch 

AN

 down to the «unlock» position. Touch 

the START/STOP pad 

1

 to start measuring.

7.  The cuff will now pump up automatically. Relax, do not move 

and do not tense your arm muscles until the measurement 

result is displayed. Breathe normally and do not talk.

8.  When the correct pressure is reached, the pumping stops and 

the pressure falls gradually. If the required pressure was not 

reached, the instrument will automatically pump some more air 

into the cuff.

9.  During the measurement, the pulse indicator 

BN

 flashes in the 

display.

10.The result, comprising the systolic 

AR

 and the diastolic 

AS

blood 

pressure and the pulse rate 

BT

 is displayed. Note also the 

explanations on further display symbols in this booklet.

11. When the measurement has finished, remove the cuff. The 

measurement values are stored automatically.

12.  Switch  off  the  device.  (The  monitor  does  switch  off 

automatically after approx. 1 min.).

How not to store a reading

Press the START/STOP pad 

1

 while the reading is being

displayed. Keep the button pressed until «M» 

BM

 is flashing and

then release it. Confirm by pressing the M-pad 

AK

 again.

•  You can stop the measurement at any time by pressing the 

START/STOP pad (e.g. if you feel uneasy or an unpleasant 

pressure sensation).

5.   Appearance of the Atrial Fibrillation Indicator for early 

Detection (only in AFIB/MAM mode)

This device is able to detect atrial fibrillation (AFIB). This

symbol 

BO

 indicates that atrial fibrillation was detected during

the measurement. If AFIB is present during blood pressure

measurement, the AFIB indicator is displayed flashing at the end

of the triple measurements. It is highly recommended to take

an additional AFIBsens

TM

 / MAM measurement an hour later to 

confirm the result. If after repeated measurement the AFIB symbol 

is no longer displayed there is no cause for concern. In such case 

it is recommended to measure again the next day.

However, if the symbol appears on a regular basis (e.g. several

times a week with measurements taken daily) we advise you to

visit your doctor. Please provide the following explanation:

Information for the doctor on frequent appearance of the

atrial fibrillation indicator

This device is an oscillometric blood pressure monitor that also

analyses pulse irregularity during measurement. The device is

clinically tested.

The AFIB symbol is displayed after the measurement, if atrial

fibrillation occurred during measuring. If the symbol appears

more frequently (e.g. several times per week on measurements

performed daily) we recommend the patient to seek medical

advice.

The device does not replace a cardiac examination, but serves

to detect atrial fibrillation that often remains undiagnosed until

stroke occurs.

 

Keep the arm still during measuring to avoid false readings.

 

This device may not detect atrial fibrillation in people with 

pacemakers or defibrillators.

6. Traffic Light Indication in the Display

The bars on the left-hand edge of the traffic light display 

BL

 show

you the range within which the indicated blood pressure value lies.

Depending on the height of the bar, the readout value is either

within the optimum (green), elevated (yellow), too high (orange) or

dangerously high (red) range. The classification corresponds to

the 4 ranges in the table as defined by the international guidelines

(ESH, AHA, JSH), as described in «Section 1.».

Содержание A7 Touch BT

Страница 1: ...EN 1 Microlife BP A7 Touch BT A7Touch BT...

Страница 2: ...Microlife BP A7 Touch BT Start Stop _...

Страница 3: ...r want to order spare parts please contact Microlife Customer Service Your dealer or pharmacy will be able to give you the address of the Microlife dealer in your country Alternatively visit the Inter...

Страница 4: ...Cleaning the cuff Accuracy test Disposal 14 Guarantee 15 Technical Specifications 1 ImportantFactsaboutBloodPressureandSelf Measurement Blood pressure is the pressure of the blood flowing in the arter...

Страница 5: ...t and pump blood Atrial fibrillation occurs when rapid disorganized electrical signals are present in the heart s two upper chambers called the atria causing them to contract quickly fast and irregula...

Страница 6: ...on The first person to measure should select user 1 Select the measuring mode standard or AFIBsensTM mode This device enables you to select either standard standard single measurement or AFIBsensTM MA...

Страница 7: ...ssure sensation 5 AppearanceoftheAtrialFibrillationIndicatorforearly Detection onlyinAFIB MAMmode This device is able to detect atrial fibrillation AFIB This symbol BO indicates that atrial fibrillati...

Страница 8: ...values 1 Firstly unlock the device then select either 1 or 2 with the user pad 9 2 Hold down the M pad AK until CL appears and then release the button 3 Touch the M pad while CL is flashing to permane...

Страница 9: ...ERR 1 Signal too weak The pulse signals on the cuff are too weak Re position the cuff and repeat the measurement ERR 2 BS Error signal During the measurement error signals were detected by the cuff c...

Страница 10: ...ections of this booklet Ensure that children do not use the instrument unsupervised some parts are small enough to be swallowed Device care Clean the instrument only with a soft dry cloth Cleaning the...

Страница 11: ...urce 4 x 1 5 V Batteries size AAA Mains adapter DC 6V 600 mA optional Battery lifetime approx 400 measurements using new batteries IP Class IP20 Reference tostandards EN 1060 1 3 4 IEC 60601 1 IEC 606...

Страница 12: ...w microlife com Asia Microlife Corporation 9F 431 RuiGuang Road NeiHu Taipei 11492 Taiwan R O C Tel 886 2 87971288 Fax 886 2 87971283 Email service microlife com tw www microlife com North Central Sou...