6 PDQeX Instruction Manual
201
90513
Load your sample into the
PDQeX extractor cartridge.
This part of the procedure will
differ depending on the sample
type. In some cases, it will be in
a suspension; in others it may be
a solid sample or a disk of
storage card.
Insert the cap into the cartridge.
Load the cartridges into the holes
of the heating block. When all
are loaded, lower the hinged flap
onto the caps of the cartridges
and close the sliding door.
MAKE SURE THE CARTRIDGE
POSITIONS
IN
THE
BLOCK
CORRESPOND
TO
THE
COLLECTION TUBES BELOW –
OTHERWISE
YOU
WILL
CONTAMINATE THE MACHINE.
WARNING; THE LOWER END OF THE PDQEX CARTRIDGE IS DESIGNED TO FIT INSIDE
THE COLLECTOR TUBE. IF IT IS ABOVE THE UPPER LIP OF THE TUBE, THERE IS A RISK
OF CROSS CONTAMINATION. THIS PROBLEM MAY OCCUR WHEN LOWER PROFILE
COLLECTION PLATES OR TUBES ARE USED.
CONTACT
MICROGEM
IF YOU NEED A COLLECTION
DRAWER TAILORED FOR THE PROFILE FOR YOUR TUBES