REI2
User manual
Doc: R2U_2_1079_002_E
Version: 1.07.9
Page
43
of
64
17.1.1. Supply
180° 5 pole socket for power supply
1 Ground
2
Not connected
3
External power supply input of from 9 to 24 allowed, maximum absoprtion 10 W
1
4
Not connected
5
Not connected
17.1.2. Analogue
I/O
240° 6 pole socket for connection of analogue acquisition devices
1
Analogue input 0 (0..5V)
2
Analogue input 1 (0..5V)
3
Analogue input 2 (0..5V)
4
5V regulated output for power supply of external devices, max 800mA
5
Ground (reference for analogue inputs)
6 Not
connected
17.1.3. Digital
I/O
240° 6 pole socket for connection of digital acquisition devices
1
Digital input/output 0 (0/5V, configurable via software)
2
Digital input/output 1 (0/5V, configurable via software)
3
Digital input/output 2 (0/5V, configurable via software)
4
Digital input/output 3 (0/5V, configurable via software)
5 Ground
6
Digital output (0/5V)
17.1.4. Radio
5 pole miniDIN socket
1 Input
2 -10V,
max
20mA
3 Output
4
Regulated 5V output, max 800mA
5 Ground
17.1.5. Computer
A
240° 6 pole socket standard signal RS 232 no hardware protocol
1 Serial
output
(TXD)
2 Not
used
3
RS232 input for flow control
4
RS232 output for flow control
5 GND
6 Serial
input
(RXD)
1
Between 9 and 11 volts REI2 is correctly powered but the internal batteries cannot be recharged.