PRECISE-LOG Data Loggers User’s Manual
Page 42/57
The valid range for these two parameters is from –32768 to 32767.
These two parameters for each channel were originally set to zero when the logger was first
released.
If you have finished the
Cali. Low
and
Cali. High
calibration instructed in the later chapter,
the “Cali. Low” and “Cali. High” values may be readjusted.
Alarm
This table specifies how each channel controls alarm state by:
Alarm Enabled:
Check this field to associate this channel to the alarm state.
Low & High Alarm:
These fields define the alarm thresholds. If the reading is beyond
these thresholds, the alarm is triggered.
WIFI Settings:
WIFI Mode:
Currently, only
Wifi Function Disabled
and
Server Mode
are available.
Wifi Function Disalbed:
disable the wifi function
Server Mode:
The logger will run as server and a client PC can
communicate with it.
Device IP/Port:
The device’s IP address is dynamically assigned by the router. You can specify the port the
device will work on.
Wifi Access Point:
This is the wifi network the logger will connect to.