MCP39F521 POWER MONITOR
DEMONSTRATION BOARD USER’S GUIDE
2015 Microchip Technology Inc.
DS50002413A-page 13
Chapter 2. Installation and Operation
2.1
GETTING STARTED
To use the MCP39F521 Power Monitor Demonstration Board, follow the steps
described in the following sections. The meter design uses a 5A load for calibration
current and a maximum current (I
MAX
) of 15A.
It is not recommended to put more than 15A through the AC plugs mounted on the
Printed Circuit Board (PCB).
To test the calibrated meter, the following connections can be made:
2.1.1
Step 1: Wiring connections
identifies the line and load connections of the MCP39F521 Power Monitor
Demonstration Board.
FIGURE 2-1:
Connecting the MCP39F521 Power Monitor Demonstration Board.
2.1.2
Step 2: Turn on line/load power to the meter (power the meter)
The meter will turn on when the line connection is powered by 90V to 220V.
2.1.3
Step 3: Connect the USB cable to a PC with the installed
“MCP39F521 Power Monitor Utility” software
Select the appropriate COM port. If the meter is connected correctly, the connection
status in the bottom-left corner of the software will display “Meter Connected”. If no
meter is found, the status will be “Meter Disconnected”. Check that the correct COM
port was selected and try again. Press the “Start” Icon to begin showing output data and
UART transmission between the PC and the MCP39F521.
LINE
LOAD
Содержание MCP39F521
Страница 1: ...2015 Microchip Technology Inc DS50002413A MCP39F521 Power Monitor Demonstration Board User s Guide...
Страница 23: ...Schematic and Layouts 2015 Microchip Technology Inc DS50002413A page 23 A 6 BOARD TOP COPPER AND SILK...
Страница 25: ...Schematic and Layouts 2015 Microchip Technology Inc DS50002413A page 25 A 8 BOARD BOTTOM COPPER...
Страница 27: ...Schematic and Layouts 2015 Microchip Technology Inc DS50002413A page 27 A 10 BOARD BOTTOM SILK...