VP-ITC MicroCalorimeter User’s Manual
1.3 Suitable Laboratory Environment
The VP-ITC with Computer Controller requires about 1 meter of normal bench space (ca. 70 cm
wide). The ThermoVac degassing station will require another 25 cm of bench space. This
location should be away from strong drafts, room temperature fluctuations, intense sunlight,
vibrations and strong electrical or magnetic fields (as may be produced by an NMR, microwave
oven, large motors or refrigeration units). In addition the mains power source (110 to 240 VAC)
should be properly grounded and free from voltage fluctuations, harmonic distortions, power dips
and spikes.
Although, the power filtering in the VP-ITC instrument is adequate for most laboratory
environments, some disturbances may affect the performance of the instrument and it may be
necessary to install a power conditioner. Since power source problems can be manifested in
many different ways, it is not possible to recommend a power conditioner for all situations. It is
recommended that you test a power conditioner, at your location, before you purchase it. If you
believe you are experiencing power source related problems, please contact a MicroCal engineer.
It is emphasized that room temperature fluctuations due to the cycling on/off of heating and
cooling systems, strong air currents, sunlight directly on the instrument and through space
electromagnetic waves may cause subtle performance problems.
4