6.Operation

1) fluorescence observation:

1. Fluorescent light source adapter 11 connect the DC power socket 3, the switch button 7 according

to "I", which is connected to power. Close the transmission light switch 9.

2 .The switch lever 5 into the fluorescent light field.

3.Specimen on the stage, will be transferred to the objective position, focusing on specimen. The

condenser is down regulated, one can make the image to achieve the most clear position.

4 With the eyepiece, adjusting the pupil distance and visibility.

5.Adjust the brightness adjustment knob 8 to achieve satisfactory lighting conditions, the default is

the maximum brightness.

6. Conversion of different magnification lens, the need to use a micro focus adjustment.

7. The weak fluorescence brightness and observed using the camera should be the eyepiece tube

switching rod 2 is completely pulled out to the 3rd gear, for the camera to capture the weak

fluorescence.

2), Bright field observation:

1. Turn off the drop shot fluorescent lighting switch button 7.

2 Bright field fluorescence switch rod 5 to the end of the module in the bright field observation stalls.

3.The right light switch 9 transmission microscope body according to the "I" side, the power supply is

switched on.

4.The lifting position 4 adjusting the condenser, the brightness knob and the aperture, in order to

achieve satisfactory lighting condition.

5 Different magnification lens, using micro focusing handwheel slightly adjusted. Observed by 4X and

10X objective, appropriate lower condenser, the lighting is more uniform. With the 100X objective,

need to put the cedar oil specimen and the objective lens in between.

Содержание MF-BG-LED

Страница 1: ...User Manual MF BG LED CX23 3T Guangzhou Micro shot Technology Co Ltd www m shot com sales m shot com...

Страница 2: ...tain it properly please read this manual carefully before operating this unit Caution This manual uses the following symbols for safety reminders Be sure to observe these warnings in order to operate...

Страница 3: ...1 CONTENT 1 Content 1 2 General note 2 3 Properties of product 2 4 Configuration 2 5 Constructure 4 6 Installation 5 7 Operation 6...

Страница 4: ...t illumination Blue EF 460nm 490nm DM 500nm BF 510nm Green EF 510nm 550nm DM 570nm BF 590nm Transmitted illumination Built in transmitted illumination system LED Power Consumption 0 5 W nominal values...

Страница 5: ...25 8 0 1 34 40X 0 65 0 6 0 52 100XO 1 25 0 13 0 27 Objective Plan 4x NA 0 10 WD 27 8 mm Plan 10x NA 0 25 WD 8 mm Plan 40x NA 0 65 WD 0 6 mm Plan 100xOil NA 1 25 WD 0 13 mm Working surrounding Indoor t...

Страница 6: ...lever 3 DC power base 7 Power button 8 Brightness adjustment knob 4 M6 fix screw 5 Lever Bright field Draw out Fluorescent Push in 6 Manual fix screw 9 Transmitted light power button 10 Transmitted l...

Страница 7: ...5 5 Installation 1 Manual fix screw M6 to fix fluorescence illumination with microscope host body 2 M6 screw to fix trinoclar head with fluorescence illumination 3 Set eyepieces...

Страница 8: ...stment 7 The weak fluorescence brightness and observed using the camera should be the eyepiece tube switching rod 2 is completely pulled out to the 3rd gear for the camera to capture the weak fluoresc...

Страница 9: ...leeve index value C said ipd 4 adjust the position of condenser and aperture Usually in the highest position using the condenser If the whole observation field of view is not enough it can reduce the...

Страница 10: ...80 4 clean the surface of the instrument can be cleaned with a soft cloth to wipe the dirt available neutral detergent scrub weight 5 custody for a long period of time without the use of a microscope...

Страница 11: ...9 Thank You...

Отзывы: