6
6.Operation
1) fluorescence observation:
1. Fluorescent light source adapter 11 connect the DC power socket 3, the switch button 7 according
to "I", which is connected to power. Close the transmission light switch 9.
2 .The switch lever 5 into the fluorescent light field.
3.Specimen on the stage, will be transferred to the objective position, focusing on specimen. The
condenser is down regulated, one can make the image to achieve the most clear position.
4 With the eyepiece, adjusting the pupil distance and visibility.
5.Adjust the brightness adjustment knob 8 to achieve satisfactory lighting conditions, the default is
the maximum brightness.
6. Conversion of different magnification lens, the need to use a micro focus adjustment.
7. The weak fluorescence brightness and observed using the camera should be the eyepiece tube
switching rod 2 is completely pulled out to the 3rd gear, for the camera to capture the weak
fluorescence.
2), Bright field observation:
1. Turn off the drop shot fluorescent lighting switch button 7.
2 Bright field fluorescence switch rod 5 to the end of the module in the bright field observation stalls.
3.The right light switch 9 transmission microscope body according to the "I" side, the power supply is
switched on.
4.The lifting position 4 adjusting the condenser, the brightness knob and the aperture, in order to
achieve satisfactory lighting condition.
5 Different magnification lens, using micro focusing handwheel slightly adjusted. Observed by 4X and
10X objective, appropriate lower condenser, the lighting is more uniform. With the 100X objective,
need to put the cedar oil specimen and the objective lens in between.