3.1.13.SUPERSPRING ZERO-
POSITIONING
51. Allow the spring to settle for at least two minutes before setting the zero-
position of the mainspring. Use the d.c. motor to move the top of the
mainspring relative to the inner support structure. The zero-position is
monitored on the front panel BNC of the Superspring controller marked
SPHERE OUT. The position can be adjusted manually or automatically to
be within about
±
20 mV of 0 V. It is preferable to set the zero-position
between 0 mV and +20 mV since long term drift will normally be
downward (negative).
The rotary knob on the front panel can be set to MANUAL, WINDOW,
AUTO, or REMOTE. The REMOTE setting should not be used. The
MANUAL setting allows the user to apply a voltage to the d.c. motor that
can be varied with a front panel mounted trimpot labeled MOTOR. The
middle of the trimpot (marked 5) applies no voltage. Numbers above the
middle (5-10) cause the motor to lift the mass (increasing positive voltage
on the sphere BNC) and numbers below the middle (0-5) lower the mass
(Decreasing the voltage on the sphere BNC).
AUTO causes the motor to seek the zero-position automatically.
!
NOTE: IN AUTO, THE DC MOTOR IS ALWAYS ACTIVE. THE
SUPERSPRING SHOULD NOT BE LEFT IN AUTO DURING NORMAL
OPERATION (WHEN THE COIL SERVO IS ACTIVE)
3-16
The WINDOW setting turns on the motor only when the spring position is
out of range (indicated by an LED on the front panel). In this case, the
AUTO mode is activated until the position is moved to zero, and then the
motor is deactivated. The Superspring can be left in this condition, but it
is still advisable to turn the rotary knob to OFF before closing the
Superspring coil loop (switch set to CLOSED on the front panel).
WINDOW mode is currently not supported by the FG5 software.
Содержание FG5
Страница 19: ...Design Components and Function 2 Figure 2 8 Top view of interferometer optics and beam path 2 11 ...
Страница 20: ...Design Components and Function 2 Figure 2 9 Side view of interferometer optics and beam path 2 12 ...
Страница 22: ...Design Components and Function 2 Figure 2 10 The Superspring 2 14 ...
Страница 31: ...Design Components and Function 2 Figure 2 13 Rotation Monitor 2 23 ...
Страница 32: ......
Страница 42: ...How to Set Up and Run the FG5 3 3 10 ...
Страница 44: ...How to Set Up and Run the FG5 3 Figure 3 2 V Post 3 12 ...
Страница 53: ...How to Set Up and Run the FG5 3 1 Backup the data 2 Shut off computer power 3 21 ...
Страница 87: ...Adjustment and Maintenance 4 4 29 ...
Страница 91: ...Adjustment and Maintenance 4 4 33 ...
Страница 104: ...Troubleshooting 5 5 2 ...
Страница 117: ...Troubleshooting 5 5 15 ...
Страница 131: ...Checklists and Logs Appendix D 9 3 ...
Страница 140: ...Checklists and Logs Appendix D Table 9 6 Replacing Drive Belt 9 12 ...
Страница 145: ...Checklists and Logs Appendix D Table 9 10 Replace Linear Bearings 9 17 ...
Страница 149: ...Checklists and Logs Appendix D Table 9 13 Replace Shaft Encoder 9 21 ...